A new method to study the efficiency of third generation blood filters
- 1 May 1993
- journal article
- Published by Wiley in British Journal of Haematology
- Vol. 84 (1) , 175-177
- https://doi.org/10.1111/j.1365-2141.1993.tb03045.x
Abstract
In order to study the efficiency of third generation blood filters we have devised a new method using 3 microns pore size polycarbonate filter membranes and a filtration chamber to trap leucocytes passing through the blood filter. The method has enabled us to make two important observations: (1) The filters function very efficiently at the start but the efficiency diminishes progressively with time. (2) When the blood flow through the filter is retarded, due to defective priming and/or filter malfunction, the filters fail even at the outset of the transfusion.Keywords
This publication has 2 references indexed in Scilit: