Failure of total condylar knee replacement. Correlation of radiographic, clinical, and surgical findings.
- 1 August 1984
- journal article
- research article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 152 (2) , 309-315
- https://doi.org/10.1148/radiology.152.2.6739790
Abstract
Fifty-five failed total condylar knee replacements are reviewed, correlating the radiographic and surgical findings at removal or revision. Failure was due to infection in 20 cases, mechanical loosening in 18, instability without loosening in 10, malalignment without loosening in 4, a supracondylar fracture in 2 and unknown causes in 1. In infection with loosening or chronic infection, radiographs usually showed a wide, extensive radiolucent region at the cement-bone interface around 1 or more components without shifting, while in acute infection without loosening the radiographs were normal. In mechanical loosening, in addition to a similar lucent region, the prosthesis became tilted in 16 out of 18 cases. Most loosening involved tilting of the tibial component into the varus position with subsidence into the medial tibial plateau, collapse of the cancellous bone and plastic deformation. Radiographs can be helpful in determining the reason for failure of a total knee replacement, which is important in planning surgery.This publication has 2 references indexed in Scilit:
- The total condylar knee prosthesis in gonarthrosis. A five to nine-year follow-up of the first one hundred consecutive replacements.Journal of Bone and Joint Surgery, 1983
- Failure of total knee arthroplasty due to loosening and deformation of the tibial component.Journal of Bone and Joint Surgery, 1978