Cramér–Rao bounds for three‐point decomposition of water and fat
Top Cited Papers
Open Access
- 9 August 2005
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 54 (3) , 625-635
- https://doi.org/10.1002/mrm.20623
Abstract
The noise analysis for three-point decomposition of water and fat was extended to account for the uncertainty in the field map. This generalization leads to a nonlinear estimation problem. The Crámer–Rao bound (CRB) was used to study the variance of the estimates of the magnitude, phase, and field map by computing the maximum effective number of signals averaged (NSA) for any choice of echo time shifts. The analysis shows that the noise properties of the reconstructed magnitude, phase, and field map depend not only on the choice of echo time shifts but also on the amount of fat and water in each voxel and their alignment at the echo. The choice of echo time shifts for spin-echo, spoiled gradient echo, and steady-state free precession imaging techniques were optimized using the CRB. The noise analysis for the magnitude explains rough interfaces seen clinically in the boundary of fat and water with source images obtained symmetrically about the spin-echo. It also provides a solution by choosing appropriate echo time shifts (−π/6 + πk, π/2 + πk, 7π/6 + πk), with k an integer. With this choice of echo time shifts it is possible to achieve the maximum NSA uniformly across all fat:water ratios. The optimization is also carried out for the estimation of phase and field map. These theoretical results were verified using Monte Carlo simulations with a newly developed nonlinear least-squares reconstruction algorithm that achieves the CRB. Magn Reson Med, 2005.Keywords
This publication has 16 references indexed in Scilit:
- Iterative decomposition of water and fat with echo asymmetry and least‐squares estimation (IDEAL): Application with fast spin‐echo imagingMagnetic Resonance in Medicine, 2005
- Referenceless PRF shift thermometryMagnetic Resonance in Medicine, 2004
- Maximum likelihood estimation of signal amplitude and noise variance from MR dataMagnetic Resonance in Medicine, 2004
- Multicoil Dixon chemical species separation with an iterative least‐squares estimation methodMagnetic Resonance in Medicine, 2003
- Optimization of diffusion measurements using Cramer‐Rao lower bound theory and its application to articular cartilageMagnetic Resonance in Medicine, 2003
- Fat‐suppressed steady‐state free precession imaging using phase detectionMagnetic Resonance in Medicine, 2003
- Is TrueFISP a gradient‐echo or a spin‐echo sequence?Magnetic Resonance in Medicine, 2003
- Noise in MRIMagnetic Resonance in Medicine, 1996
- Separation of fat and water in fast spin‐echo MR imaging with the three‐point dixon techniqueJournal of Magnetic Resonance Imaging, 1995
- Three‐point dixon technique for true water/fat decomposition with B0 inhomogeneity correctionMagnetic Resonance in Medicine, 1991