Clinical comparison of frequency doubling technology perimetry and Humphrey perimetry
Open Access
- 1 March 2001
- journal article
- clinical trial
- Published by BMJ in British Journal of Ophthalmology
- Vol. 85 (3) , 360-362
- https://doi.org/10.1136/bjo.85.3.360
Abstract
AIM To determine the number of missed points on frequency doubling technology (FDT) perimetry that optimise the sensitivity and specificity of the test and to determine the topographical accuracy of the test in a clinical setting. METHODS In a prospective study, the perimetric data from 99 patients who underwent both FDT perimetry in the screening mode and Humphrey 24-2 (H24-2) were used to determine the sensitivity and specificity of the FDT perimetry compared with the full threshold H24-2 as the gold standard. RESULTS Missed points on the FDT perimetry correlated with both the mean deviation and the corrected pattern standard deviation on the Humphrey perimetry. A score assigned to abnormal points on the FDT perimetry and the Humphrey total deviation plot showed a significant correlation for both the location and the depth of the defect. In comparing the Humphrey hemifield test with the FDT perimetry results, if at least one missed point on the frequency doubling test was considered as abnormal then the overall sensitivity of the test was 78.1% and the specificity was 89.1%. CONCLUSION FDT perimetry in the screening mode performed in a clinical setting was highly specific, exhibited reasonable sensitivity, and accurately determined the location and depth of scotomas when compared with the full threshold Humphrey 24-2.Keywords
This publication has 13 references indexed in Scilit:
- Comparison of frequency doubling perimetry with Humphrey visual field analysis in a glaucoma practiceAmerican Journal of Ophthalmology, 2000
- Algorithm for interpreting the results of frequency doubling perimetryAmerican Journal of Ophthalmology, 2000
- Frequency doubling technology perimetry for detection of glaucomatous visual field lossAmerican Journal of Ophthalmology, 2000
- Screening for glaucoma with frequency-doubling technology and Damato campimetry.Archives of Ophthalmology (1950), 1999
- Testing for glaucoma with the frequency‐doubling illusion in the whole, macular and eccentric visual fieldsAustralian and New Zealand Journal of Ophthalmology, 1999
- Frequency doubling perimetry in glaucoma early diagnosisActa Ophthalmologica Scandinavica, 1998
- Identification of glaucoma-related visual field abnormality with the screening protocol of frequency doubling technologyAmerican Journal of Ophthalmology, 1998
- Clinical classification of glaucomatous visual field loss by frequency doubling perimetryAmerican Journal of Ophthalmology, 1998
- Differential vulnerability of neurochemically identified subpopulations of retinal neurons in a monkey model of glaucomaBrain Research, 1995
- Chronic Human Glaucoma Causing Selectively Greater Loss of Large Optic Nerve FibersOphthalmology, 1988