LXXXI. A high precision pulse height analyser of moderately high speed
- 29 July 1951
- journal article
- research article
- Published by Taylor & Francis in Journal of Computers in Education
- Vol. 42 (330) , 792-806
- https://doi.org/10.1080/14786445108561308
Abstract
A pulse height analyser is described suited particularly to the requirements of nuclear physics. Information is stored in a supersonic delay line and displayed on a cathode ray tube. The accuracy of channel width and the linearity of response are both one part in a thousand. The number of channels and storage capacity per channel may be varied at will to give 60 channels of up to 106 pulses, 80 channels of up to 3 × 104 pulses, or 120 channels of up to 1023 pulses. The maximum counting rate is 1600 per second.Keywords
This publication has 8 references indexed in Scilit:
- LXXX. A design of an ultrasonic delay-lineJournal of Computers in Education, 1951
- A single channel pulse analyser for nuclear experimentsJournal of Scientific Instruments, 1951
- A stable ninety-nine channel pulse amplitude analyser for slow countingMathematical Proceedings of the Cambridge Philosophical Society, 1950
- The Ten-Channel Electrostatic Pulse AnalyzerReview of Scientific Instruments, 1949
- A Pulse Amplitude Analyzer for Nuclear Research Using Pretreated PulsesReview of Scientific Instruments, 1949
- A Pulse Analyser for Nuclear ResearchReview of Scientific Instruments, 1947
- Analyseur d'amplitude automatique par numération pour chambre d'ionisation à amplificateur proportionnelJournal de Physique et le Radium, 1946
- A Differential Pulse Amplitude SelectorReview of Scientific Instruments, 1940