Comparison of computed aortic blood velocity with that of electromagnetic flowmeter
- 1 May 1962
- journal article
- research article
- Published by American Physiological Society in Journal of Applied Physiology
- Vol. 17 (3) , 482-486
- https://doi.org/10.1152/jappl.1962.17.3.482
Abstract
The probe of an electromagnetic flowmeter was placed on the ascending aorta of five dogs. Aortic blood pressure was obtained from the same area. This pressure signal was fed into a simple computer circuit designed to solve a modification of the Navier-Stokes equation. The output of this circuit was assumed to be proportional to blood velocity and was compared directly with the velocity obtained with the electromagnetic flowmeter. The peak amplitudes and the stroke volumes (represented by the area under the velocity tracings), obtained simultaneously by the two different methods, were compared. Each dog was bled, transfused, and given methoxamine and nitroglyc-erin to produce a wide variation in stroke volume and blood pressure. A correlation coefficient of .93 was obtained from 35[degree] paired stroke -volume values. Thus, it appears that in a wide variety of conditions, stroke volume may be derived from the aortic pressure pulse by the computer technique.Keywords
This publication has 4 references indexed in Scilit:
- Relationship of aortic pressure and diameter in the dogJournal of Applied Physiology, 1961
- VELOCITY OF BLOOD FLOW AND STROKE VOLUME OBTAINED FROM THE PRESSURE PULSE *Journal of Clinical Investigation, 1959
- An Electric Device for Instantaneous and Continuous Computation of Aortic Blood VelocityCirculation Research, 1957
- A Square Wave Electromagnetic Flowmeter for Application to Intact Blood VesselsCirculation Research, 1955