Is it necessary to transport arterial blood samples on ice for pH and gas analysis?
- 1 September 1984
- journal article
- research article
- Published by Springer Nature in Canadian Journal of Anesthesia/Journal canadien d'anesthésie
- Vol. 31 (5) , 568-571
- https://doi.org/10.1007/bf03009545
Abstract
We evaluated whether arterial blood samples for pH and blood gas analysis need to be transported on ice. We found that although the changes in pH, pCO2 and pO2 were greater in samples kept at room temperature versus those kepi on ice, the difference was probably not of clinical significance until the period of time after arterial puncture exceeded 20 minutes. We recommend that arterial blood samples do not need to be kept on ice if the analysis for pH and gases is performed within 20 minutes of blood being drawn. Les échantiilons pour gaiométrie doivent-ils étre transpones dans un baquet de glace? Nous nous sommes posés cede question et avons trouvé que même si les valeurs de pH, de pCO2 et de pO2 étaienl plus élevées dans les echantiilons gardés à la température de la pièce, les différences observées sont probablement sans significa-tionclinique à condition que I’analyse seffectue moins de 20 minutes apreès le prélèvement. Nous croyons done qu ’ it n’est pas nécessaire de garder ces échantiilons sur la glace si l’analyse peut s’effectuer en moins de 20 minutes après le prélèvement.Keywords
This publication has 5 references indexed in Scilit:
- Blood gas analysis: effect of air bubbles in syringe and delay in estimation.BMJ, 1982
- Air bubbles and temperature effect on blood gas analysis.Journal of Clinical Pathology, 1980
- Pseudohypoxemia Secondary to Leukemia and ThrombocytosisNew England Journal of Medicine, 1979
- The effects of air bubbles and time delay on blood gas analysis.1974
- Leakage of Oxygen from Blood and Water Samples Stored in Plastic and Glass SyringesBMJ, 1971