Salient characteristics of the electron capture detector (ECD) are reviewed in the light of insights and inconsistencies found in the current literature. The first part of this paper focuses on conditions influencing response, and notes a few recent advances in detector technology and operation. In the second part, we illustrate some of the principles discussed with temperature-programmed separations of chlorinated hydrocarbons.