MEASUREMENT OF GLOMERULAR-FILTRATION RATE WITH TC-99M DTPA - COMPARISON OF PLASMA-CLEARANCE TECHNIQUES
- 1 March 1987
- journal article
- research article
- Vol. 28 (3) , 372-377
Abstract
We have compared several techniques for measuring [99mTc]DTPA plasma clearance following single injection to estimate glomerular filtration rate (GFR). Half hourly measurements of disappearance of plasma activity were used to calculate a reference GFR corrected for one-pool assumption and body surface area. Alternative methods involving (i) single blood sample, (ii) two blood samples, (iii) external detector clearance rate, and (iv) a combination of (i) and (iii) were then compared. Closest correlations were obtained with (i) two blood samples at 2 hr and 4 hr (s.e.e. 2.8 ml/min) and (i.v.) external rate constant from 2-5 hr with a blood sample at 3 hr (s.e.e. 3.0 ml/min). Correlations with single blood sample were closest at 3 hr and 4 hr postinjection (s.e.e. 5.4 and 4.5 ml/min, respectively). External detector disappearance rate constant alone was least accurate (s.e.e. > 10 ml, min).This publication has 2 references indexed in Scilit:
- Simultaneous Measurement of Renal Clearance and Plasma Clearance of 99mTc-Labelled Diethylenetriaminepenta-Acetate, 51Cr-Labelled Ethylenediaminetetra-Acetate and Inulin in ManClinical Science, 1984
- COMPARISON OF 4 COMMERCIAL TC-99M(SN)DTPA PREPARATIONS USED FOR THE MEASUREMENT OF GLOMERULAR-FILTRATION RATE - CONCISE COMMUNICATION1980