The Prosthesis-Bone Interface Adjacent to Tibial Components Inserted without Cement. Clinical and Radiographic Follow-up at Nine to Twelve Years*
- 1 March 1996
- journal article
- Published by Wolters Kluwer Health in Journal of Bone and Joint Surgery
- Vol. 78 (3) , 340-7
- https://doi.org/10.2106/00004623-199603000-00004
Abstract
1983, and none were lost to follow-up. When they were last seen, thirty-four (92 per cent) of the thirty-seven knees had a well fixed tibial component, both clinically and radiographically.Eight patients (eight knees; 22 per cent) died. Eleven patients (eleven knees; 30 per cent) had a revision at an average of sixty-five months (range, four to ninety-five months) postoperatively. The reason for the revision was failure of a metal-backed patellar component in three knees, excessive wear of the polyethylene of the tibial component in one, a hematogenous infection in four, aseptic loosening of the tibial component in two, and chronic synovitis in one. The remaining seventeen patients (eighteen knees; 49 per cent) were seen for clinical and radiographic follow-up at an average of eleven years (range, nine to twelve years) after the operation.With removal of the implant as the end point, the cumulative rate of survival was 83 per cent at fifty-six months and 67 per cent at 108 months. Some of the failures were secondary to features of the prosthetic design that currently are considered to be inadequate, including a metal-backed patellar component and carbon- fiber-reinforced polyethylene. In the twenty-six knees in which the prosthesis had been retained, the implant was stable and the prosthesis-bone interface was unchanged as seen radiographically at the time of the most recent follow-up examination. This finding demonstrates that a porous ingrowth surface is capable of providing a secure interface for biological fixation over the long term....Keywords
This publication has 32 references indexed in Scilit:
- Cemented and Ingrowth Fixation of the Miller???Galante ProsthesisClinical Orthopaedics and Related Research, 1990
- Results of Revision Total Knee Arthroplasty Performed for Aseptic LooseningPublished by Wolters Kluwer Health ,1990
- Mechanisms of Failure of the Femoral and Tibial Components in Total Knee ArthroplastyPublished by Wolters Kluwer Health ,1989
- Patellar Component Failure in Cementless Total Knee ArthroplastyPublished by Wolters Kluwer Health ,1988
- Failure Mechanisms of Metal-Backed Patellar ComponentsPublished by Wolters Kluwer Health ,1988
- Further Observations on Metal-Backed Patellar Component FailureClinical Orthopaedics and Related Research, 1988
- The Total Condylar ProsthesisClinical Orthopaedics and Related Research, 1986
- Preliminary Experience with a Total Knee Prosthesis with Porous Coating Used without CementPublished by Wolters Kluwer Health ,1983
- Powder Metal-Made Orthopedic Implants with Porous Surface for Fixation by Tissue IngrowthClinical Orthopaedics and Related Research, 1983
- Nonparametric Estimation from Incomplete ObservationsJournal of the American Statistical Association, 1958