Design Consideration for a Continuously Rotating Positron Computed Tomograph
- 1 January 1982
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 29 (1) , 534-538
- https://doi.org/10.1109/TNS.1982.4335902
Abstract
Design consideration for a whole-body, multislice Positron CT having 3 detector rings, 160 BGO crystal/ring, and a continuously rotating scan is described. The one detector ring has 40 "BGO Quad-detectors", each of which consists of 4 BGO crystals and two photomultipliers. Improvements in the BGO crystal and the photomultiplier have been accomplished. Geometrical configuration of the collimator was determined through optimization of the main performances of both the inplane and cross-plane images, utilizing the formulae for a cylindrical phantom with uniform activity. A continuously rotating scan (0.5-1rps) with unequally spaced the Quad-detector units is used. Along with this scan, a 32-channel rotary photo-coupler was developed for contact-free data transmission.Keywords
This publication has 14 references indexed in Scilit:
- Electronics syteem for positron computerized tomography “positologica” dedicated to head studiesNuclear Instruments and Methods in Physics Research, 1982
- Recent developments in photomultipliers for nuclear radiation detectorsNuclear Instruments and Methods in Physics Research, 1982
- Analytical Study of the Performance of a Multilayer Positron Computed Tomography ScannerJournal of Computer Assisted Tomography, 1982
- A quad BGO detector and its timing and positioning discrimination for positron computed tomographyNuclear Instruments and Methods in Physics Research, 1982
- IMAGING PROPERTIES OF A POSITRON TOMOGRAPH WITH 280 BGO CRYSTALSJournal of Computer Assisted Tomography, 1981
- Design of the Neuro-ECAT®: A High-Resolution, High Efficiency Positron Tomograph for Imaging the Adult Head or Infant TorsoIEEE Transactions on Nuclear Science, 1981
- Positologica: A Positron ECT Device with a Continuously Rotating Detector RingIEEE Transactions on Nuclear Science, 1980
- Design of a High Resolution Positron Emission TomographJournal of Computer Assisted Tomography, 1980
- Positome II: A High Efficiency Positron Imaging Device for Dynamic Brain StudiesIEEE Transactions on Nuclear Science, 1979
- A Multislice Positron Emission Computed Tomograph (PETT IV) Yielding Transverse and Longitudinal ImagesRadiology, 1978