Reliability Indexes of Automated Perimetric Tests
- 1 September 1988
- journal article
- research article
- Published by American Medical Association (AMA) in Archives of Ophthalmology (1950)
- Vol. 106 (9) , 1252-1254
- https://doi.org/10.1001/archopht.1988.01060140412043
Abstract
• The "reliability" of a subject's automated perimetric test result is generally assessed by three measures: fixation loss and false-positive and false-negative rates. These reliability criteria were examined for 76 glaucomatous and 248 normal subjects who underwent visual field testing (C-30-2 program; Humphrey Visual Field Analyzer, Allergan Humphrey, San Leandro, Calif). Of the examination results, 45% in glaucomatous subjects and 30% in normal controls were considered unreliable with the use of the manufacturer's reliability criteria. Most test results were unreliable because they failed to meet the criterion for fixation loss. The greater rejection rate among glaucomatous subjects was entirely due to their higher rate of false-negative responses. Factors such as age, pupil diameter, and visual acuity did not explain the difference between the false-negative rates of glaucomatous patients and normal subjects.This publication has 6 references indexed in Scilit:
- Recognizing Glaucomatous Field Loss With the Humphrey STATPACArchives of Ophthalmology (1950), 1987
- Evaluation of Nerve Fiber Layer AssessmentArchives of Ophthalmology (1950), 1984
- Differential Light Threshold in Automated Static PerimetryArchives of Ophthalmology (1950), 1984
- Differential Light ThresholdArchives of Ophthalmology (1950), 1984
- Variability of Static Visual Threshold Responses in Patients With Elevated IOPsArchives of Ophthalmology (1950), 1982
- Early Visual Field Disturbances in GlaucomaArchives of Ophthalmology (1950), 1977