Optimizing the tensile properties of polyvinyl alcohol hydrogel for the construction of a bioprosthetic heart valve stent
- 1 January 2002
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 63 (6) , 854-861
- https://doi.org/10.1002/jbm.10333
Abstract
Although bioprosthetic heart valves offer the benefits of a natural opening and closing, better hemodynamics, and avoidance of life-long anticoagulant therapy, they nevertheless tend to fail in 10–15 years from tears and calcification. Several authors, including the present ones, have identified the rigid stent as a factor contributing to these failures. The ultimate solution is an artificial heart valve that has mechanical properties that allow it to move in conformity with the aortic root during the cardiac cycle, has superior hemodynamics, is nonthrombogenic, will last more than 20 years, and mitigates the need for anticoagulants. We have identified a polymer, polyvinyl alcohol (PVA) hydrogel, that has mechanical properties similar to soft tissue. The purpose of this research is to match the tensile properties of PVA to the porcine aortic root and to fabricate a stent prototype for a bioprosthetic heart valve with the use of the PVA hydrogel. Specimens of 15% w/w PVA were prepared by processing through 1–6 cycles of freezing (−20 °C) at 0.2 °C/min freeze rate and thawing (+20°C) at different thawing rates (0.2 °C/min and 1 °C/min), for different holding times (1 and 6 h) at −20 °C. Subsequently tensile tests and stress-relaxation tests were conducted on the specimens. The different holding times at −20 °C demonstrated no difference in the result. The slower thawing rate improved the tensile properties but did not produce significant changes on the stress-relaxation properties. The nonlinear stress–strain curve for the PVA after the fourth freeze–thaw cycle matched the porcine aortic root within the physiological pressure range. The stress-relaxation curve for PVA also approximated the shape of the aortic root. The complex geometry of an artificial heart valve stent was successfully injection molded. These results, in combination with other preliminary findings for biocompatibility and fatigue behavior, suggest that PVA hydrogel is a promising biomaterial for implants, catheters, and artificial skin. © 2002 Wiley Periodicals, Inc. J Biomed Mater Res (Appl Biomater) 63: 854–861, 2002Keywords
This publication has 28 references indexed in Scilit:
- An Optimal Mounting Frame to Reduce Flexural Stresses of Bioprosthetic Heart ValvesAsaio Journal, 1994
- Geometry of homograft valve leaflets: Effect of dilation of the aorta and the aortic rootThe Annals of Thoracic Surgery, 1993
- A computerized system for video analysis of the aortic valveIEEE Transactions on Biomedical Engineering, 1990
- Effects of age on the anisotropy of the descending human thoracic aorta determined by uniaxial tensile testing and digestion by NaOH under loadCanadian Journal of Physiology and Pharmacology, 1989
- Comparative study of primary tissue valve failure between lonescu-shiley pericardial and Hancock porcine valves in the aortic positionThe American Journal of Cardiology, 1986
- Long-term failure rate and morphologic correlations in porcine bioprosthetic heart valvesThe American Journal of Cardiology, 1983
- Creep and Flexibility Measurements on Plastic Frames for Tissue Heart ValvesThe Thoracic and Cardiovascular Surgeon, 1981
- The cyclic changes and structure of the base of the aortic valveAmerican Heart Journal, 1980
- Poly(vinyl alcohol) hydrogels for synthetic articular cartilage materialJournal of Biomedical Materials Research, 1973
- Tricuspid Leaflet Aortic Valve ProsthesisCirculation, 1966