An evaluation of the mini-Wright peak flow meter.
- 1 February 1979
- Vol. 34 (1) , 79-81
- https://doi.org/10.1136/thx.34.1.79
Abstract
The mini-Wright peak flow meter (MPFM) was evaluated. A strong positive correlation (r = 0.970) with the Wright''s peak flow meter (PFM) was shown. MPFM measurements were biased to be about 38 l/min higher than PFM measurements (95% confidence limits 31.0-45.0 l/min). Between instrument variation was found (F-ratio 3.67 with 9-81 degrees of freedom: P < 0.001). In practice this did not appreciably affect individual measurements greatly as 95% confidence limits on any individual measurements were increased from .+-. 24 l/min to .+-. 27 l/min. There was no significant day-to-day variability in measurements obtained with individual instruments. The MPFM is a pocket-sized, simple, cheap and robust instrument for following changes in human ventilatory function. In clinical trials and surveys, both the bias in favor of the MPFM compared to the PFM and inter-machine variation must be taken into account. As the manufacturers have altered the scale to remove the bias since this study was performed, it will be important to know whether the original or the modified meter is being used in future studies.This publication has 6 references indexed in Scilit: