Reduction of patient doses from barium meal and barium enema examinations through changes in equipment factors
- 1 December 1994
- journal article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 67 (804) , 1196-1205
- https://doi.org/10.1259/0007-1285-67-804-1196
Abstract
A study of patient doses for barium meal and barium enema examinations has been made for X-ray departments throughout Grampian Region. Dose-area products are substantially less than UK reference levels. Contributions from fluoroscopy and radiography varied significantly between different units, with fluoroscopy making up 35-92% of the mean doses for barium meals and 24-57% for barium enemas. Equipment related factors had a greater influence on patient doses than radiologists' techniques. The gain of the image intensifier and the exposure factors selected by the automatic exposure control (AEC) were the most important factors determining doses for fluoroscopy. Changes in kV/mA factors selected by the AEC have allowed reductions of 20-50% to be made in effective dose from fluoroscopy with some units. The method used for recording images was the major influence on radiographic doses. Digital spot images or fluorographic films taken from the intensifier gave only 10-20% of the dose with a film/screen system and are recommended where the image quality is satisfactory for the application. Changes in film/screen systems have given worthwhile reductions in radiographic doses. If all factors are optimized, mean doses for barium examinations could be reduced to 20% of current reference levels.Keywords
This publication has 10 references indexed in Scilit:
- A study of the relationship between patient dose and size in paediatric radiologyThe British Journal of Radiology, 1994
- Implementation of a programme for reduction of radiographic doses and results achieved through increases in tube potentialThe British Journal of Radiology, 1993
- Estimation of effective dose to the patient during medical X-ray examinations from measurements of the dose-area productPhysics in Medicine & Biology, 1992
- Audit in action: significant reduction of double-contrast barium enema screening time with no loss of examination qualityThe British Journal of Radiology, 1992
- Minimizing radiation dose to patient and staff during fluoroscopic, nasoenteral tube insertionsThe British Journal of Radiology, 1992
- Effective dose equivalents, HE, in diagnostic radiologyMedical Physics, 1990
- Dependence of the photon backscatter factor for water on source-to-phantom distance and irradiation field sizePhysics in Medicine & Biology, 1990
- The influence of tube filtration and potential on patient dose during X-ray examinationsPhysics in Medicine & Biology, 1988
- An investigation into the levels of radiation exposure in diagnostic examinations involving fluoroscopyThe British Journal of Radiology, 1987
- Radiation dose received by patients during barium meal investigations under automatic brightness controlThe British Journal of Radiology, 1985