Guidance to users of Nycomed Amersham and North American Scientific, Inc., I-125 Interstitial Sources: Dosimetry and calibration changes: Recommendations of the American Association of Physicists in Medicine Radiation Therapy Committee Ad Hoc Subcommittee
- 9 April 1999
- journal article
- guideline
- Published by Wiley in Medical Physics
- Vol. 26 (4) , 570-573
- https://doi.org/10.1118/1.598570
Abstract
Dose calculations to patients undergoing implantation of 125I interstitial brachytherapy sources are affected by two recent changes in low-energy seed dosimetry: (a) implantation of a new primary air-kerma strength standard at the National Institute of Standards and Technology (NIST) on 1 January 1999 and (b) publication of revised dose-rate distributions in AAPM's Task Group 43 Report. The guidance herein represents AAPM's recommendations for users of 125I interstitial seed products marketed prior to 1 January 1999 (Nycomed Amersham models 6711 and 6702 and North American Scientific, Inc. models 3631 A/S and 3631 A/M. Implementation of Task Group 43 (TG43) 125I dose calculations involves revising data stored in files of radiation treatment planning software and lowering the prescribed dose to be delivered to patients by as much as 15% to avoid modifying the dose actually delivered to patients. The magnitude of the dose prescription change depends on the dosimetry data used prior to TG43 and the implant geometry. Adapting to the revised NIST calibration standard requires the user to increase the dose-rate constant (or its equivalent by 11.5%) but does not require modification of the prescribed dose. Failure to correctly implement these modifications can result in 20% or even 30% errors.Keywords
This publication has 8 references indexed in Scilit:
- Evaluation of a new brachytherapy iodine‐125 source by AAPM TG43 formalismMedical Physics, 1998
- Clinical Impact of Implementing the Recommendations of AAPM Task Group 43 on Permanent Prostate Brachytherapy Using 125IInternational Journal of Radiation Oncology*Biology*Physics, 1998
- Report of the ad hoc committee of the AAPM Radiation Therapy Committee on 125I sealed source dosimetryInternational Journal of Radiation Oncology*Biology*Physics, 1998
- Code of practice for brachytherapy physics: Report of the AAPM Radiation Therapy Committee Task Group No. 56Medical Physics, 1997
- Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No. 43Medical Physics, 1995
- Monte Carlo evaluation of specific dose constants in water for 1 2 5 I seedsMedical Physics, 1988
- Exposure Standardization of Iodine-125 Seeds Used for BrachytherapyJournal of Research of the National Bureau of Standards, 1984
- Physical dosimetry of 125I seeds of a new design for interstitial implantInternational Journal of Radiation Oncology*Biology*Physics, 1983