The Pipeline Embolization Device for the Intracranial Treatment of Aneurysms Trial
Top Cited Papers
Open Access
- 9 December 2010
- journal article
- clinical trial
- Published by American Society of Neuroradiology (ASNR) in American Journal of Neuroradiology
- Vol. 32 (1) , 34-40
- https://doi.org/10.3174/ajnr.a2421
Abstract
Endoluminal reconstruction with flow diverting devices represents a novel constructive technique for the treatment of cerebral aneurysms. We present the results of the first prospective multicenter trial of a flow-diverting construct for the treatment of intracranial aneurysms. Patients with unruptured aneurysms that were wide-necked (> 4 mm), had unfavorable dome/neck ratios (<1.5), or had failed previous therapy were enrolled in the PITA trial between January and May 2007 at 4 (3 European and 1 South American) centers. Aneurysms were treated with the PED with or without adjunctive coil embolization. All patients underwent clinical evaluation at 30 and 180 days and conventional angiography 180 days after treatment. Angiographic results were adjudicated by an experienced neuroradiologist at a nonparticipating site. Thirty-one patients with 31 intracranial aneurysms (6 men; 42-76 years of age; average age, 54.6 years) were treated during the study period. Twenty-eight aneurysms arose from the ICA (5 cavernous, 15 parophthalmic, 4 superior hypophyseal, and 4 posterior communicating segments), 1 from the MCA, 1 from the vertebral artery, and 1 from the vertebrobasilar junction. Mean aneurysm size was 11.5 mm, and mean neck size was 5.8 mm. Twelve (38.7%) aneurysms had failed (or recurred after) a previous endovascular treatment. PED placement was technically successful in 30 of 31 patients (96.8%). Most aneurysms were treated with either 1 (n = 18) or 2 (n = 11) PEDs. Fifteen aneurysms (48.4%) were treated with a PED alone, while 16 were treated with both PED and embolization coils. Two patients experienced major periprocedural stroke. Follow-up angiography demonstrated complete aneurysm occlusion in 28 (93.3%) of the 30 patients who underwent angiographic follow-up. No significant in-construct stenosis (≥ 50%) was identified at follow-up angiography. Intracranial aneurysm treatment with the PED is technically feasible and can be achieved with a safety profile analogous to that reported for stent-supported coil embolization. PED treatment elicited a very high rate (93%) of complete angiographic occlusion at 6 months in a population of the most challenging anatomic subtypes of cerebral aneurysms.Keywords
This publication has 23 references indexed in Scilit:
- Treatment of Intracranial Aneurysms by Functional Reconstruction of the Parent Artery: The Budapest Experience with the Pipeline Embolization DeviceAmerican Journal of Neuroradiology, 2010
- Coiling of Intracranial AneurysmsStroke, 2009
- Clinical experience with Matrix2 360° coils in the treatment of 100 intracranial aneurysmsSurgical Neurology, 2009
- Curative cerebrovascular reconstruction with the Pipeline embolization device: the emergence of definitive endovascular therapy for intracranial aneurysmsJournal of NeuroInterventional Surgery, 2009
- CURATIVE ENDOVASCULAR RECONSTRUCTION OF CEREBRAL ANEURYSMS WITH THE PIPELINE EMBOLIZATION DEVICENeurosurgery, 2009
- CURATIVE RECONSTRUCTION OF A GIANT MIDBASILAR TRUNK ANEURYSM WITH THE PIPELINE EMBOLIZATION DEVICENeurosurgery, 2009
- Endovascular Treatment of Large and Giant AneurysmsAmerican Journal of Neuroradiology, 2008
- Durability of Aneurysm Embolization with Matrix Detachable CoilsNeurosurgery, 2006
- Long-Term Angiographic Recurrences After Selective Endovascular Treatment of Aneurysms With Detachable CoilsStroke, 2003
- Endovascular Treatment of Experimental Aneurysms with Liquid Polymers: The Protective Potential of StentsNeurosurgery, 1996