Glutathione and cysteine measurement in biological samples by HPLC with a glassy carbon working detector
- 1 May 1994
- journal article
- research article
- Published by Wiley in Biomedical Chromatography
- Vol. 8 (3) , 134-136
- https://doi.org/10.1002/bmc.1130080308
Abstract
An HPLC method using a glassy carbon working detector for measuring cysteine and reduced glutathione (GSH) in biological samples was developed. Oxidized glutathione (GSSG) was also measured after reduction with glutathione reductase. This study was conducted in human plasma. Cysteine and GSH standard curves were linear in the physiological range, presenting detection limits of 22.0 pmol and 6.0 pmol respectively. Plasmatic results found were 43.92 ± 4.15, 4.50 ± 0.65, and 0.19 ± 0.12 mM (means ± SE), for cysteine, GSH and GSSG, respectively. Peak specificity for cysteine and reduced glutathione was certified by their disappearance after treatment with N‐ethylmaleimide.Keywords
This publication has 21 references indexed in Scilit:
- Determination of glutathione and glutathione disulfide using glutathione reductase and 2-vinylpyridinePublished by Elsevier ,2004
- Chloride secretory response to extracellular ATP in human normal and cystic fibrosis nasal epitheliaAmerican Journal of Physiology-Cell Physiology, 1992
- Separation of reduced and oxidized glutathione in a pharmaceutical preparation by ion-interaction reversed-phase HPLCJournal of Pharmaceutical and Biomedical Analysis, 1992
- Influence of neuronal uptake on pre- and postjunctional effects of ?-adrenoceptor agonists in tissues with noradrenaline ? ATP cotransmissionNaunyn-Schmiedebergs Archiv für experimentelle Pathologie und Pharmakologie, 1991
- The hepatic gluthathione system — influences of xenobioticsExperimental pathology, 1990
- Determination of glutathione in biological tissues by high-performance liquid chromatography with electrochemical detectionJournal of Pharmaceutical and Biomedical Analysis, 1989
- Measurement of Tissue Sulfhydryls and Disulfides in Tissue Protein and Nonprotein Fractions by High Performance Liquid Chromatography Using Electrochemical DetectionJournal of Liquid Chromatography, 1987
- Reduced and oxidized glutathione efflux from liverFEBS Letters, 1978
- A fluorometric method for determination of oxidized and reduced glutathione in tissuesAnalytical Biochemistry, 1976
- Tissue sulfhydryl groupsArchives of Biochemistry and Biophysics, 1959