Four‐dimensional computed tomography: Image formation and clinical protocol
Top Cited Papers
- 16 March 2005
- journal article
- Published by Wiley in Medical Physics
- Vol. 32 (4) , 874-889
- https://doi.org/10.1118/1.1869852
Abstract
Respiratory motion can introduce significant errors in radiotherapy. Conventional CT scans as commonly used for treatment planning can include severe motion artifacts that result from interplay effects between the advancing scan plane and object motion. To explicitly include organ/target motion in treatment planning and delivery, time-resolved CT data acquisition (4D Computed Tomography) is needed. 4DCT can be accomplished by oversampled CT data acquisition at each slice. During several CT tube rotations projection data are collected in axial cine mode for the duration of the patient's respiratory cycle (plus the time needed for a full CT gantry rotation). Multiple images are then reconstructed per slice that are evenly distributed over the acquisition time. Each of these images represents a different anatomical state during a respiratory cycle. After data acquisition at one couch position is completed, x rays are turned off and the couch advances to begin data acquisition again until full coverage of the scan length has been obtained. Concurrent to CT data acquisition the patient's abdominal surface motion is recorded in precise temporal correlation. To obtain CT volumes at different respiratory states, reconstructed images are sorted into different spatio-temporally coherent volumes based on respiratory phase as obtained from the patient's surface motion. During binning, phase tolerances are chosen to obtain complete volumetric information since images at different couch positions are reconstructed at different respiratory phases. We describe 4DCT image formation and associated experiments that characterize the properties of 4DCT. Residual motion artifacts remain due to partial projection effects. Temporal coherence within resorted 4DCT volumes is dominated by the number of reconstructed images per slice. The more images are reconstructed, the smaller phase tolerances can be for retrospective sorting. From phantom studies a precision of about 2.5 mm for quasiregular motion and typical respiratory periods could be concluded. A protocol for 4DCT scanning was evaluated and clinically implemented at the MGH. Patient data are presented to elucidate how additional patient specific parameters can impact 4DCT imaging.Keywords
This publication has 15 references indexed in Scilit:
- Comparison of helical and cine acquisitions for 4D-CT imaging with multislice CTMedical Physics, 2005
- Internal target volume determined with expansion margins beyond composite gross tumor volume in three-dimensional conformal radiotherapy for lung cancerInternational Journal of Radiation Oncology*Biology*Physics, 2004
- Acquiring 4D thoracic CT scans using a multislice helical methodPhysics in Medicine & Biology, 2004
- 4D-CT imaging of a volume influenced by respiratory motion on multi-slice CTMedical Physics, 2004
- A method for the reconstruction of four‐dimensional synchronized CT scans acquired during free breathingMedical Physics, 2003
- Respiration‐correlated spiral CT: A method of measuring respiratory‐induced anatomic motion for radiation treatment planningMedical Physics, 2002
- Acquiring a four-dimensional computed tomography dataset using an external respiratory signalPhysics in Medicine & Biology, 2002
- Multiple “slow” CT scans for incorporating lung tumor mobility in radiotheraphy planningInternational Journal of Radiation Oncology*Biology*Physics, 2001
- Motion Artifacts in Subsecond Conventional CT and Electron-Beam CT: Pictorial Demonstration of Temporal ResolutionRadioGraphics, 2000
- Analysis of movement of intrathoracic neoplasms using ultrafast computerized tomographyInternational Journal of Radiation Oncology*Biology*Physics, 1990