In Vitro Performance Testing of Three Small Volume-Holding Chambers under Conditions That Correspond with Use by Infants and Small Children
- 1 January 1997
- journal article
- research article
- Published by Mary Ann Liebert Inc in Journal of Aerosol Medicine
- Vol. 10 (4) , 341-349
- https://doi.org/10.1089/jam.1997.10.341
Abstract
The in vitro behavior of three types of small volume-holding chambers intended for use by infants and small children (child AeroChamber®, Vent-170 and Space Chamber™) has been compared with two pMDI-delivered aerosol formulations (salbutamol: 100 μg unit dose and beclomethasone di-propionate (BDP): 50 μg unit dose) widely prescribed for this age group of patients. All devices were evaluated using a pediatric breathing simulator that created respiratory conditions likely to be encountered with infants and small children. The filter that collected aerosol delivered from the holding chamber on test was located at approximately the position that the patient's lips would be in the mask, by means of an annular support plate glued into the mask itself. At the lowest tidal volume (50 ml), no salbutamol or BDP was delivered by either the Vent-170 or Space Chamber™, whereas the unit doses from the child AeroChamber® were 39.7 ± 1.6 μg and 11.8 ± 1.2 μg, respectively. The Vent-170 and Space Chamber™ delivered measurable doses of both drugs when the tidal volume was increased to 100 ml and again to 200 ml, however, the corresponding doses available from the AeroChamber® were always significantly greater. These findings emphasize the importance of designing in vitro tests that mimic use by the patient group for which the device is intended. In vitro measurements made at constant high flow rates in excess of 20 l/min do not reveal these differences in performance that are clinically significant, and may lead the physician to prescribe a device that under certain conditions may not deliver any drug to infants or small children.Keywords
This publication has 6 references indexed in Scilit:
- Particle Size Measurements for the Space-Chamber Metered Dose Inhaler Holding Chamber Compared with Aerochamber and Metered Dose Inhaler AloneJournal of Aerosol Medicine, 1997
- Effect of electrostatic charge, flow, delay and multiple actuations on the in vitro delivery of salbutamol from different small volume spacers for infants.Thorax, 1996
- The effect of handling techniques on electrostatic charge on spacer devices: A correlation with in vitro particle size analysisInternational Journal of Pharmaceutics, 1996
- The Optimum Size and Shape of Spacer Devices for Inhalational TherapyJournal of Aerosol Medicine, 1995
- Improvement in sodium cromoglycate delivery from a spacer device by use of an antistatic lining, immediate inhalation, and avoiding multiple actuations of drug.Thorax, 1993
- Design and characteristics of a portable breath actuated, particle size selective medical aerosol inhalerJournal of Aerosol Science, 1982