Prosthetic Aortic Stenosis
- 1 May 1974
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation
- Vol. 49 (5) , 921-924
- https://doi.org/10.1161/01.cir.49.5.921
Abstract
A cause of early death after aortic valve replacement with a caged-ball prosthesis is obstruction to left ventricular outflow because the prosthesis is too large for the aortic root. Of 68 patients dying within two months of aortic valve replacement, death in ten, each of whom had had intractable low cardiac output after operation, was attributed at necropsy to prosthetic aortic stenosis, despite the use of small sized (8A Starr-Edwards) prostheses in seven of them. The diameters of the aorta at the sinotubular junction, determined from the preoperative cineangiograms, in the seven patients with prosthetic stenosis were < 30 mm in all. Poppet clearances, defined as the differences between poppet and aortic root diameters, ranged from 4 to 12 mm (avg. 9). In contrast, the diameters of the aortas at the sinotubular junctions in eight control patients (unobstructed prosthetic aortic valves and early death from other causes) were > 30 in all but one, and the poppet clearances ranged from 12 to 19 mm (avg. 15). Thus, prosthetic aortic stenosis is likely to develop after aortic valve replacement with rigid-framed caged ball valves if the preoperative aortograms disclose aortic diameters at the sinotubular junctions to be < 30 mm. In such patients, either the aorta must be widened for a caged-ball prosthesis or a central flow valve must be used.Keywords
This publication has 10 references indexed in Scilit:
- Causes of Death and other Anatomic Observations after Cardiac Valve ReplacementPublished by S. Karger AG ,2015
- Pathologic anatomy of cardiac valve replacement: A study of 224 necropsy patientsProgress in Cardiovascular Diseases, 1973
- Late hemodynamic studies after mitral valve surgeryThe Journal of Thoracic and Cardiovascular Surgery, 1973
- Hemodynamic confirmation of obstruction to left ventricular inflow by a caged-ball prosthetic mitral valveThe Journal of Thoracic and Cardiovascular Surgery, 1973
- Hemodynamic studies in patients with cloth-covered composite-seat Starr-Edwards valve prosthesesThe Journal of Thoracic and Cardiovascular Surgery, 1970
- Causes of early postoperative death following cardiac valve replacementThe Journal of Thoracic and Cardiovascular Surgery, 1967
- Prosthetic Replacement of the Mitral ValveCirculation, 1967
- Mechanisms of acute left atrial thrombosis after mitral valve replacement: Pathologic findings indicating obstruction to left atrial emptyingThe American Journal of Cardiology, 1966