Vector implementation of Chang's attenuation correction method for single photon emission computed tomography
- 1 November 1987
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 14 (6) , 1045-1047
- https://doi.org/10.1118/1.595983
Abstract
A vector version of Chang''s attenuation correction method for single photon emission computed tomography accommodating arbitrary convex body contours is presented and its implementation on a general purpose array processor described. For a constant linear attenuation coefficient .mu. = 0.10 cm-1 and contour shapes similar to those encountered in clinical work, correction factors calculated with the vector algorithm are the same to within a few percent as those obtained using the original Chang method. Typical execution speed of the new algorithm is 0.01 s/correction factor for a 64 point contour.This publication has 3 references indexed in Scilit: