Localized 1H NMR spectroscopy of rat spinal cord in Vivo
Open Access
- 1 April 1996
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 35 (4) , 443-448
- https://doi.org/10.1002/mrm.1910350401
Abstract
A movable, actively decoupled surface coil has been employed to obtain a localized 1H NMR spectrum from the lumbosacral spinal cord of a live Lewis rat. A volume selective ‘VOSY’ normally spelled out as ‘volume selective spectroscopy’ spectroscopy pulse sequence that incorporates “phase ramped” selective RF pulses, has been used to minimize random phase jitter in the NMR signal as a result of the large frequency shifts required to locate the voxel in the center of the cord while using intense gradient pulses. Spectra from 13-μl voxels in healthy rats and in rats inoculated with guinea pig spinal cord and complete Freund's adjuvant, resulting in experimental autoimmune encephalomyelitis, are shown.Keywords
Funding Information
- Centre for Magnetic Resonance of the University of Queensland
This publication has 17 references indexed in Scilit:
- Magnetic resonance imaging of antibody-mediated demyelinating experimental allergic encephalomyelitisJournal of Neuroimmunology, 1994
- Quantitative MRI of spinal cord injury in a rat modelMagnetic Resonance in Medicine, 1994
- Magnetic resonance spectroscopy of multiple sclerosis: in-vivo detection of myelin breakdown productsPublished by Elsevier ,1993
- 1H NMR properties of N‐acetylaspartylglutamate in extracts of nervous tissue of the ratNMR in Biomedicine, 1992
- The effects of RF phase shifts and instabilities caused by carrier frequency switching in localized NMR spectroscopyJournal of Magnetic Resonance (1969), 1990
- Localized proton spectroscopy using stimulated echoesJournal of Magnetic Resonance (1969), 1987
- Volume-selective multipulse spin-echo spectroscopyJournal of Magnetic Resonance (1969), 1987
- Selected Volume Excitation Using Stimulated Echoes (VEST). Applications to spatially localized spectroscopy and imagingJournal of Magnetic Resonance (1969), 1986
- Water signal elimination in Vivo, using “suppression by mistimed echo and repetitive gradient episodes”Journal of Magnetic Resonance (1969), 1986
- Image-selected in Vivo spectroscopy (ISIS). A new technique for spatially selective nmr spectroscopyJournal of Magnetic Resonance (1969), 1986