Simultaneous Quantitation of Prostaglandins F 1? . (PGF 1? ) and PGF 2aL by Selected Ion Recording with a Deuterated Internal Standard
- 1 January 1978
- journal article
- research article
- Published by Taylor & Francis in Analytical Letters
- Vol. 11 (3) , 195-205
- https://doi.org/10.1080/00032717808059693
Abstract
A gas chromatographic-mass spectrometric method for the simultaneous measurement of PGF1α, and PGF2α at the ng level is reported. The internal standard was [3, 3, 4, 4-D] PGF2α. The method involves conversion of the substrates to their methyl ester-trimethyl-silyl ether (ME-TMS) derivatives and selected ion recording. The standard curves were reproducible and linear between 0. 5 and 40 ng (1. 4 to 112 picomole). The regression characteristics of the standard curves for each PGF were virtually identical to those of the analogous curves from separate determinations.Keywords
This publication has 6 references indexed in Scilit:
- Gas Liquid Chromatographic—Mass Spectrometric Methods for Quantitation of Prostaglandin AnalogsAnalytical Letters, 1976
- A rapid gas chromatographic-mass spectrometric method for prostaglandin analysis at picomole levelsAnalytical Biochemistry, 1974
- Modified prostaglandins as internal standards for quantitative analysis of prostaglandins of the E series by combined gas chromatography-mass spectrometryProstaglandins, 1973
- Methods for quantitative analysis of PGF2α, PGE2, 9α,11α-dihydroxy-15-keto-prost-5-enoic acid and 9α,11α,15-trihydroxy-prost-5-enoic acid from body fluids using deuterated carriers and gas chromatography-mass spectrometryAnalytical Biochemistry, 1973
- Quantitative determination of prostaglandins A, B and E in the sub-nanogram rangeProstaglandins, 1973
- Mass spectrometric determination of picomole amounts of prostaglandins E2 and F2α using synthetic deuterium labeled carriersBiochemical and Biophysical Research Communications, 1971