Radiochromic film dosimetry of a low energy proton beam
- 11 July 2000
- journal article
- Published by Wiley in Medical Physics
- Vol. 27 (7) , 1655-1660
- https://doi.org/10.1118/1.599032
Abstract
In this work some dosimetric characteristics of MD-55-2 GafChromic films were studied in a low energy proton beam (21.5 MeV) directly in a water phantom. The nonlinearity of the optical density was quantified by a factor P(lin). A correction factor P(en), that accounts for optical density dependence on the energy, was empirically determined. The effects of detector thickness in depth dose measurements and of the film orientation with respect to beam direction were investigated. The results show that the MD-55-2 films provide dose measurements with the films positioned perpendicularly to the proton beam. A dosimetric formalizm is proposed to determine the dose to water at depth d, with films oriented perpendicularly to the beam axis. This formalism uses a calibration factor of the radiochromic film determined directly on the proton beam at a reference depth in water, and the P(lin) factor, that takes into account the nonlinearity of the calibration curve and the P(en) factor that, in turn takes into account the change of proton beam energy in water. The MD-55-2 films with their high spatial resolution and the quasiwater equivalent material are attractive, positioned perpendicularly along the beam axis, for the absolute dose determination of very small beam sizes and modulated proton beams.Keywords
This publication has 8 references indexed in Scilit:
- Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55Medical Physics, 1998
- Radiochromic film dosimetry for clinical proton beamsApplied Radiation and Isotopes, 1997
- Quantitative evaluation of radiochromic film response for two‐dimensional dosimetryMedical Physics, 1997
- Dosimetric characteristics of an improved radiochromic filmMedical Physics, 1996
- Preliminary Assessment of LiF and Alanine Detectors for the Dosimetry of Proton Therapy BeamsRadiation Protection Dosimetry, 1996
- Calculated Bragg curves for ionization chambers of different shapesMedical Physics, 1995
- Ionization chamber dosimetry of proton beams using cylindrical and plane parallel chambers. Nwversus NKion chamber calibrationsPhysics in Medicine & Biology, 1995
- Supplement to the code of practice for clinical proton dosimetryRadiotherapy and Oncology, 1994