The NASPE */BPEG ** Defibrillator Code
- 1 September 1993
- journal article
- guideline
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 16 (9) , 1776-1780
- https://doi.org/10.1111/j.1540-8159.1993.tb01809.x
Abstract
A new generic code, patterned after and compatible with the NASPE/BPEG Generic Pacemaker Code (NBG Code) was adopted by the NASPE Board of Trustees on January 23, 1993. It was developed by the NASPE Mode Code Committee, including members of the North American Society of Pacing and Electrophysiology (NASPE) and the British Pacing and Electrophysiology Group (BPEG). It is abbreviated as the NBD (for NASPE/BPEG Defibrillator) Code. It is intended for describing the capabilities and operation of implanted cardioverter defibrillators (ICDs) in conversation, record keeping, and device labeling, and incorporates four positions designating: (1) shock location; (2) antitachycardia pacing location; (3) means of tachycardia detection; and (4) antibradycardia pacing location. An additional Short Form, intended only for use in conversation, was defined as a concise means of distinguishing devices capable of shock alone, shock plus antibradycardia pacing, and shock plus antitachycardia and antibradycardia pacing.Keywords
This publication has 2 references indexed in Scilit:
- The NASPE*/BPEG** Generic Pacemaker Code for Antibradyarrhythmia and Adaptive‐Rate Pacing and Antitachyarrhythmia DevicesPacing and Clinical Electrophysiology, 1987
- Implantable Cardiac Pacemakers Status Report and Resource GuidelineCirculation, 1974