Effect of Blood Collection Tubes on Total Triiodothyronine and Other Laboratory Assays
- 1 February 2005
- journal article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 51 (2) , 424-433
- https://doi.org/10.1373/clinchem.2004.043349
Abstract
Background: Increased total triiodothyronine (TT3) assay results in apparently euthyroid patients triggered an investigation of the effect of blood collection tubes on serum TT3 and other laboratory assays. Methods: We examined potential assay interference for three types of tubes: plastic Greiner Bio-One™ Vacuette™; glass Becton Dickinson (BD) Vacutainer™; and plastic BD Vacutainer SST™ tubes. Serum samples from apparently healthy volunteers (age range, 30–60 years; 15 males and 34 females) were collected in different tube types and analyzed in 17 immunoassays (n = 49), 30 clinical chemistry tests (n = 20), and 33 immunology assays (n = 15). Tube effects were also examined by adding pooled serum to different tube types. Results: TT3 values, when measured by the IMMULITE™ 2000 but not the AxSYM™ analyzer, were significantly higher (P <0.0001) for SST (2.81 nmol/L) than either glass (2.15 nmol/L) or Vacuette (2.24 nmol/L) tubes. The effect was large enough to substantially shift the distribution of patient values, increasing the percentage of values above the reference interval from 11.3% to 35.8%. The degree of interference from SST tubes on TT3 differed among various tube lots and could be attributed to a tube additive shared by other plastic tubes. Results from several other tests statistically differed among tube types, but differences were not considered to be clinically significant. Conclusions: Assay interferences from blood collection tubes represent challenges to clinical laboratories because they are not detected by the usual quality-control or proficiency testing programs. Laboratories can, however, address this problem by monitoring distribution of patients’ results.Keywords
This publication has 30 references indexed in Scilit:
- Potential Interferences from Blood Collection Tubes in Mass Spectrometric Analyses of Serum PolypeptidesClinical Chemistry, 2004
- Evaluation of a new generation of plastic evacuated blood-collection tubes in clinical chemistry, therapeutic drug monitoring, hormone and trace metal analysiscclm, 2004
- Influence of blood collection in plastic vs. glass evacuated serum-separator tubes on hormone and tumour marker levelscclm, 2004
- Stability of Therapeutic Drugs in Serum Collected in Vacutainer Serum Separator Tubes Containing a New Gel (SST II)Therapeutic Drug Monitoring, 2001
- Comparison of BD Vacutainer SST™ Plus Tubes with BD SST™ II Plus Tubes for common analytesClinica Chimica Acta; International Journal of Clinical Chemistry, 2001
- Suitability of Collection Tubes with Separator Gels for Collecting and Storing Blood Samples for Therapeutic Drug Monitoring (TDM)cclm, 2000
- Stability of Therapeutic Drug Measurement in Specimens Collected in VACUTAINER Plastic Blood-Collection TubesTherapeutic Drug Monitoring, 1996
- A new plastic evacuated tube with plasma separatorJournal of Clinical Laboratory Analysis, 1995
- Absorption of Therapeutic Drugs by Barrier Gels in Serum Separator Blood Collection Tubes:Volume- and Time-dependent Reduction in Total and Free Drug ConcentrationsaAmerican Journal of Clinical Pathology, 1994
- Suitability of Collection Tubes with Separator Gels for Therapeutic Drug MonitoringTherapeutic Drug Monitoring, 1990