Accuracy of Pulse Oximetry During Arterial Oxyhemoglobin Desaturation in Dogs
Open Access
- 1 January 1988
- journal article
- research article
- Published by Wolters Kluwer Health in Anesthesiology
- Vol. 68 (1) , 111-114
- https://doi.org/10.1097/00000542-198801000-00018
Abstract
An animal model was developed to evaluate the accuracy of pulse oximetry over a wide range of oxyhemoglobin desaturation. The fractional inspired oxygen concentration was varied from 0.03-1.0 in five anesthetized dogs. One hundred and twelve simultaneous pulse oximeter oxygen saturation measurements (SpO2) and IL 282 CO-Oximeter arterial oxygen saturation (SaO2) measurements were made. Variance of SpO2 was increased for SaO2 < 22%. Linear regression analysis of the data for SaO2 > 22% produced the equation y = 0.93x + 9.8 (r2 = 0.97). The mean difference between SpO2 and SaO2 was +5.5% .+-. 4.2% (SD) over the range of 22-100%. Spectral analyses of oxygenated (O2Hb) and reduced (RHb) canine and human hemoglobins were performed. The absorption spectra of canine O2Hb and RHb were nearly identical to those of human O2Hb and RHb. Therefore, 1) SpO2 measurements in dogs at SaO2 > 22% are relatively accurate, and 2) hemoglobin absorption characteristics support the contention that such canine pulse oximeter studies can be extrapolated to humans.This publication has 4 references indexed in Scilit:
- Range of Accuracy of Two Wavelength OximetryChest, 1986
- Continuous Monitoring of Arterial Oxygen Saturation with Pulse Oximetry during Transfer to the Recovery RoomAnesthesia & Analgesia, 1985
- Pulse oximetry in pediatric intensive care: Comparison with measured saturations and transcutaneous oxygen tensionThe Journal of Pediatrics, 1985
- Measuring percent oxygen saturation of hemoglobin, percent carboxyhemoglobin and methemoglobin, and concentrations of total hemoglobin and oxygen in blood of man, dog, and baboon.Clinical Chemistry, 1980