Calculation of the absorbed dose distribution due to irregularly shaped photon beams using pencil beam kernels derived from basic beam data
- 1 April 1996
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 41 (4) , 637-656
- https://doi.org/10.1088/0031-9155/41/4/005
Abstract
In radiotherapy, accurately calculated dose distributions of irregularly shaped photon beams are needed. In this paper, an algorithm is presented which enables the calculation of dose distributions due to irregular fields using pencil beam kernels derived from simple basic beam data usually measured on treatment units, i.e. central axis depth - dose curves and profiles. The only extra data that are needed, and are not currently measured, is the phantom scatter factor curve at the reference depth. The algorithm has been developed as an extension to a previously developed algorithm for rectangular fields which is based on the Milan - Bentley storage model. In the case of an irregular field, the depth dose and the boundary function are computed by convolution of a field intensity function with pencil beam kernels. The depth dose is computed by using a `scatter' kernel, which is derived from the stored depth - dose curves and from the phantom scatter factor curve. The boundary function is computed by using a `boundary' kernel, which is derived from the boundary profile of a number of large square fields. Because of the simplicity of the data used and the underlying concepts, which for instance do not separate the head scatter from the primary beam, this algorithm presents some shortcomings. On the other hand, this simplicity is also of great advantage and the inaccuracy is acceptable for most clinical situations.Keywords
This publication has 14 references indexed in Scilit:
- A table of phantom scatter factors of photon beams as a function of the quality index and field sizePhysics in Medicine & Biology, 1996
- A pencil beam model for photon dose calculationMedical Physics, 1992
- Collapsed cone convolution of radiant energy for photon dose calculation in heterogeneous mediaMedical Physics, 1989
- Extraction of pencil beam kernels by the deconvolution methodMedical Physics, 1988
- Use of fast Fourier transforms in calculating dose distributions for irregularly shaped fields for three‐dimensional treatment planningMedical Physics, 1987
- A convolution method of calculating dose for 15‐MV x raysMedical Physics, 1985
- A photon dose distribution model employing convolution calculationsMedical Physics, 1985
- Revision of tissue‐maximum ratio and scatter‐maximum ratio concepts for cobalt 60 and higher energy x‐ray beamsMedical Physics, 1980
- The storage and manipulation of radiation dose data in a small digital computerThe British Journal of Radiology, 1974
- The Extension of the Concept of Tissue-Air Ratios (TAR) to High-Energy X-Ray BeamsRadiology, 1970