Studies on bromobenzene‐induced hepatotoxicity using in vivo MR microscopy with surgically implanted RF coils
- 1 June 1994
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 31 (6) , 619-627
- https://doi.org/10.1002/mrm.1910310607
Abstract
Using surgically implanted RF coils at 300 MHz, three-dimensional microscopic MR images of rat liver were obtained in vivo to follow the development of pathology induced by bromobenzene exposure. Formalin fixed specimens of liver from these animals were also imaged using in vitro MR microscopy, followed by conventional optical microscopy. All MR images were acquired using a spin-warp pulse sequence with TR = 950 ms and TE= 23 ms. The in vivo images were reconstructed as 2562 × 32 arrays with a voxel size of (50 μm)2 × 219 pm, while the in vitro images were reconstructed as 2562 × 128 arrays, giving an isotropic resolution at (39 μm)2. Based on results from six animals, we have found in all animals exposed to bromobenzene, image intensity decreased in specific hepatic tissue regions. These regions were well correlated to low signal intensity areas observed in in vitro MR images at higher resolution. Conventional optical microscopy indicated that the low signal intensity regions corresponded to areas of necrosis. The decrease in signal intensity is consistent with increased local diffusion coefficients as a result of necrosis. This study demonstrates that MR microscopy with implanted RF coils can be successfully used to follow tissue pathological changes in living tissues.Keywords
This publication has 26 references indexed in Scilit:
- Implanted coil MR microscopy of renal pathologyMagnetic Resonance in Medicine, 1989
- Amplitude- and frequency/phase-modulated refocusing pulses that induce plane rotations even in the presence of inhomogeneous B1 fieldsJournal of Magnetic Resonance (1969), 1988
- Studies of diffusion in random fields produced by variations in susceptibilityJournal of Magnetic Resonance (1969), 1988
- Inductive coupling and tuning in NMR probes; ApplicationsJournal of Magnetic Resonance (1969), 1988
- The contribution of bromobenzene to our current understanding of chemically-induced toxicitiesLife Sciences, 1988
- In Situ Magnetic Resonance MicroscopyInvestigative Radiology, 1987
- Wireless implanted magnetic resonance probes for in vivo NMRJournal of Magnetic Resonance (1969), 1986
- Nuclear Magnetic Resonance Spectroscopy and Imaging in the Study of Experimental Liver DiseasesAlcohol, Clinical and Experimental Research, 1986
- A Ventilator for Magnetic Resonance ImagingInvestigative Radiology, 1986
- An in Vivo NMR probe circuit for improved sensitivityJournal of Magnetic Resonance (1969), 1983