Quality assurance of the dose delivered by small radiation segments
- 1 September 1998
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 43 (9) , 2665-2675
- https://doi.org/10.1088/0031-9155/43/9/017
Abstract
The use of intensity modulation with multiple static fields has been suggested by many authors as a way to achieve highly conformal fields in radiotherapy. However, quality assurance of linear accelerators is generally done only for beam segments of 100 MU or higher, and by measuring beam profiles once the beam has stabilized. We propose a set of measurements to check the stability of dose delivery in small segments, and present measured data from three radiotherapy centres. The dose delivered per monitor unit, MU, was measured for various numbers of MU segments. The field flatness and symmetry were measured using either photographic films that are subsequently scanned by a densitometer, or by using a diode array. We performed the set of measurements at the three radiotherapy centres on a set of five different Philips SL accelerators with energies of 6 MV, 8 MV, 10 MV and 18 MV. The dose per monitor unit over the range of 1 to 100 MU was found to be accurate to within % of the nominal dose per monitor unit as defined for the delivery of 100 MU for all the energies. For four out of the five accelerators the dose per monitor unit over the same range was even found to be accurate to within %. The flatness and symmetry were in some cases found to be larger for small segments by a maximum of 9% of the flatness/symmetry for large segments. The result of this study provides the dosimetric evidence that the delivery of small segment doses as top-up fields for beam intensity modulation is feasible. However, it should be stressed that linear accelerators have different characteristics for the delivery of small segments, hence this type of measurement should be performed for each machine before the delivery of small dose segments is approved. In some cases it may be advisable to use a low pulse repetition frequency (PRF) to obtain more accurate dose delivery of small segments.Keywords
This publication has 12 references indexed in Scilit:
- Intensity-modulated radiotherapy by means of static tomotherapy: A planning and verification studyMedical Physics, 1997
- Dosimetric evaluation of compensation in radiotherapy of the breast: MLC intensity modulation and physical compensatorsRadiotherapy and Oncology, 1997
- Number and orientations of beams in intensity‐modulated radiation treatmentsMedical Physics, 1997
- Output stability of a linear accelerator during the first three secondsMedical Physics, 1996
- Design of compensators for breast radiotherapy using electronic portal imagingRadiotherapy and Oncology, 1995
- Which is the most suitable number of photon beam portals in coplanar radiation therapy?International Journal of Radiation Oncology*Biology*Physics, 1995
- Dosimetric problems at low monitor unit settings for scanned and scattering foil electron beamsMedical Physics, 1994
- X-ray field compensation with multileaf collimatorsInternational Journal of Radiation Oncology*Biology*Physics, 1994
- Dosimetric accuracy at low monitor unit settingsThe British Journal of Radiology, 1991
- Teletherapy beam characteristics: The first secondMedical Physics, 1987