Incidence, Time Course, and Characteristics of Microbubble Formation During Radiofrequency Ablation of Pulmonary Veins with an 8‐mm Ablation Catheter
- 14 September 2006
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 29 (9) , 979-984
- https://doi.org/10.1111/j.1540-8159.2006.00473.x
Abstract
Microbubble formation during pulmonary vein (PV) radiofrequency (RF) ablation of atrial fibrillation (AF) occurs relatively frequently. Prior studies have shown that microbubble formation may be associated with an increased risk of complications. However, the incidence, time course, and temperature characteristics of microbubble formation during AF ablation with an 8-mm catheter have not been prospectively described in humans.We studied 46 (30 men, age 56+/-10 years) patients with AF who underwent RF ablation of PVs between January 2005 and December 2005 using an 8F, 8-mm Biosensetrade mark ablation catheter (Biosense-Webster, Diamond Bar, CA, USA). All patients underwent continuous intracardiac echocardiography (ICE). Microbubble patterns were classified as either type 1 (intermittent, scattered microbubble formation) or type 2 (explosive shower of dense microbubbles). Formation of any microbubbles was detected by ICE and the time, PV location, and electrode temperature were recorded.A total of 1,479 (32+/-13, range 12-73) RF lesions were delivered to 167 veins. Twenty (2%) lesions were classified as type 2. Since the number of lesions resulting in type 2 bubbles was very small, only type 1 lesions were included in the final analysis. Thirty-nine (85%) patients had at least one lesion associated with bubble formation during ablation (mean: 7+/-7 lesions, range 1-28 lesions). Twenty-three percent (327) of the RF lesions resulted in bubble formation. RF generator power setting during lesions resulting in bubble formation was lower than lesions which did not result in bubble formation (47.9+/-7.4 W vs 49.7+/-7.1 W, P<0.001). Logistic regression analysis revealed a significant negative correlation (P<0.001) between RF generator power settings and a positive correlation between the generator temperature settings and formation of bubbles (both P<0.02). However, the maximum temperature attained was not different between lesions resulting in bubble formation (n=327) and those which did not result in bubble formation (n=1,139). Fifty-three (16%) of the lesions associated with bubble formation occurred within 2-10 seconds after RF was begun. Bubble formation was significantly more frequent in left superior PVs compared to the other PVs (left superior PV 27.3% left inferior PV 18.6%, right superior PV 20.5%, and right inferior PV 18.8%, P=0.005, left superior PV vs other PVs, P<0.001) even after adjustment for the other factors including generator power settings and the temperature setting.Bubble formation is common during RF ablation of PV with 8-mm tip catheter and can occur as early as 2 seconds after starting RF. RF generator power is negatively correlated with bubble formation while generator temperature settings are positively correlated with formation of bubbles. Microbubble formation is also more frequent with ablation of the left superior PV probably due to better catheter contact in that area.Keywords
This publication has 13 references indexed in Scilit:
- Avoiding Microbubbles Formation During Radiofrequency Left Atrial Ablation Versus Continuous Microbubbles Formation and Standard Radiofrequency Ablation Protocols: Comparison of Energy Profiles and Chronic Lesion CharacteristicsJournal of Cardiovascular Electrophysiology, 2006
- Percutaneous Pulmonary Vein Stenting for the Treatment of Severe Stenosis After Pulmonary Vein IsolationJournal of Cardiovascular Electrophysiology, 2005
- Assessment of Temperature, Proximity, and Course of the Esophagus During Radiofrequency Ablation Within the Left AtriumCirculation, 2005
- Microbubbles during radiofrequency catheter ablation: Composition and formationHeart Rhythm, 2005
- Worldwide Survey on the Methods, Efficacy, and Safety of Catheter Ablation for Human Atrial FibrillationCirculation, 2005
- Analysis of Catheter-Tip (8-mm) and Actual Tissue Temperatures Achieved During Radiofrequency Ablation at the Orifice of the Pulmonary VeinCirculation, 2004
- Atrio-Esophageal Fistula as a Complication of Percutaneous Transcatheter Ablation of Atrial FibrillationCirculation, 2004
- Pulmonary Vein Stenosis After Radiofrequency Ablation of Atrial FibrillationCirculation, 2003
- Phased-Array Intracardiac Echocardiography Monitoring During Pulmonary Vein Isolation in Patients With Atrial FibrillationCirculation, 2003
- Effect of Heating on Pulmonary Veins:Journal of Cardiovascular Electrophysiology, 2003