Significant reduction in minute ventilation and peak inspiratory pressures with arteriovenous CO sub 2 removal during severe respiratory failure
- 1 April 1997
- journal article
- research article
- Published by Wolters Kluwer Health in Critical Care Medicine
- Vol. 25 (4) , 689-695
- https://doi.org/10.1097/00003246-199704000-00022
Abstract
To quantify CO2 removal using an extracorporeal low-resistance membrane gas exchanger placed in an arteriovenous shunt and evaluate its effects on the reduction of ventilatory volumes and airway pressures during severe respiratory failure induced by smoke inhalation injury. Prospective study. Research laboratory. Adult female sheep (n = 5). Animals were instrumented with femoral and pulmonary arterial catheters and underwent an LD50 cotton smoke inhalation injury via a tracheostomy under halothane anesthesia. Twenty-four hours after smoke inhalation injury, the animals were reanesthetized and systemically heparinized for cannulation of the left carotid artery and common jugular vein to construct a simple arteriovenous shunt. A membrane gas exchanger was interposed within the arteriovenous shunt, and blood flow produced by the arteriovenous pressure gradient was unrestricted at the time of complete recovery from anesthesia. CO2 removal by the gas exchanger was measured as the product of the sweep gas flow (FIO2 of 1.0 at 2.5 to 3.0 L/min) and the exhaust CO2 content measured with an inline capnometer. CO2 removed by the animal's lungs was determined by the expired gas CO2 content in a Douglas bag. We made stepwise, 20% reductions in ventilator support hourly. We first reduced the tidal volume to achieve a peak inspiratory pressure of 100 torr (>13.3 kPa) at maximally reduced ventilator support. Mean arterial pressure and cardiac output did not change significantly as a result of arteriovenous shunting. Extracorporeal CO2 removal using a low-resistance gas exchanger in a simple arteriovenous shunt allows significant reduction in minute ventilation and peak inspiratory pressure without hypercapnia or the complex circuitry and monitoring required for conventional extracorporeal membrane oxygenation. Arteriovenous CO2 removal can be applied as an easy and cost-effective treatment to minimize ventilator-induced barotrauma and volutrauma during severe respiratory failure. (Crit Care Med 1997; 25:689-695)Keywords
This publication has 28 references indexed in Scilit:
- Intracorporeal gas exchange: Current status and future developmentIntensive Care Medicine, 1996
- Strategies to Reduce Surface Area Requirements for Carbon Dioxide Removal for an Intravenacaval Gas Exchange DeviceAsaio Journal, 1995
- Effect of Il-1 Blockade on Inflammatory Manifestations of Acute Ventilator-Induced Lung Injury in a Rabbit ModelExperimental Lung Research, 1995
- Performance of an intravenous gas exchanger (IVOX) in a venovenous bypass circuitThe Annals of Thoracic Surgery, 1994
- Use of computational fluid dynamics in the design of the Avecor Affinity oxygenatorPerfusion, 1994
- Randomized, prospective trial of pressure-limited versus volume-controlled ventilation in severe respiratory failureCritical Care Medicine, 1994
- Femoral arteriovenous extracorporeal carbon dioxide elimination using low blood flowCritical Care Medicine, 1992
- Prolonged pumpless arteriovenous perfusion for carbon dioxide extractionThe Annals of Thoracic Surgery, 1991
- Ventilatory management of ARDS: Can it affect the outcome?Intensive Care Medicine, 1990
- Low-Frequency Positive Pressure Ventilation with Extracorporeal Carbon Dioxide Removal (LFPPV-ECCO2R)Anesthesia & Analgesia, 1978