Abstract
Debris from polyethylene wear causes osteolysis. In this study, we examined the effect of acetabular liner modularity on polyethylene wear and osteolysis. We compared forty-one hips (thirty-nine patients) treated with a nonmodular, porous-coated acetabular component with a matched group of forty-one hips (forty patients) treated with a modular acetabular component. The groups were matched by patient gender and age, type of polyethylene material, method of polyethylene sterilization, femoral head size and manufacturer, and stem manufacturer. The mean follow-up period was 5.3 years (range, 3.8 to 6.8 years) for the nonmodular group and 5.5 years (range, 3.8 to 8.0 years) for the modular group. Using serial radiographs and a computer-assisted method, we measured two-dimensional head penetration into the polyethylene liner. Temporal head-penetration data and linear regression analysis were used to calculate the true wear rates. The nonmodular acetabular components demonstrated a lower, but not a significantly lower, mean true wear rate than did the modular components (0.11 compared with 0.16 mm/yr, p = 0.22), and they were associated with a significantly lower rate of osteolysis (2% compared with 22%, p = 0.01). In addition, the true wear rates of the nonmodular components were less variable than those of the modular components. The 95% confidence interval for the wear rates of the nonmodular components (0.08 to 0.13 mm/yr) was nearly half that of the modular group (0.11 to 0.20 mm/yr). The lower and more consistent true wear rates of the nonmodular components could be attributed to the fact that these cups were designed to have greater liner-shell conformity, greater liner thickness, and less liner-shell micromotion than modular components. These design factors could have favorably altered the stress distribution throughout the liner and could have thereby decreased wear. Although nonmodular components may present a partial solution to the problems of wear and osteolysis, they pose a disadvantage when a failed liner in a bone-ingrown acetabular component needs to be revised.Keywords
This publication has 17 references indexed in Scilit:
- Factors contributing to rapid wear and osteolysis in hips with modular acetabular bearings made of HylamerThe Journal of Arthroplasty, 2000
- Two-Dimensional Versus Three-Dimensional Radiographic Measurements of Polyethylene WearPublished by Wolters Kluwer Health ,1999
- Determination of Polyethylene Wear in Total Hip Replacements with Use of Digital Radiographs*Journal of Bone and Joint Surgery, 1997
- Radiographic Evaluation of Penetration by the Femoral Head into the Polyethylene Liner over Time*Journal of Bone and Joint Surgery, 1997
- The role of backside polishing, cup angle, and polyethylene thickness on the contact stresses in metal-backed acetabular componentsJournal of Biomechanics, 1997
- Fractures through cystic lesions of the greater trochanter: A cause of late pain after cementless total hip arthroplastyThe Journal of Arthroplasty, 1996
- Increasing incidence of femoral osteolysis in association with uncemented Harris—Galante total hip arthroplasty: A follow-up reportThe Journal of Arthroplasty, 1996
- Patterns of osteolysis around total hip components inserted with and without cement.Journal of Bone and Joint Surgery, 1995
- Severe osteolysis of the pelvic in association with acetabular replacement without cement.Journal of Bone and Joint Surgery, 1993