Multi-Channel Mass-Separated Neutral Particle Energy Analyser for Simultaneous Measurements of Hydrogen and Deuterium Atoms Emitted from Tokamak Plasma
- 1 November 1983
- journal article
- Published by IOP Publishing in Japanese Journal of Applied Physics
- Vol. 22 (11R)
- https://doi.org/10.1143/jjap.22.1709
Abstract
A mass-separated neutral particle energy analyser able to carry out the simultaneous measurement of hydrogen and deuterium atoms emitted from a tokamak plasma was constructed. The analyser was calibrated for the energy and mass separation in the energy range from 0.4 keV to 9 keV. It was installed in the JFT-2 tokamak in order to investigate the behavior of deuterons and protons in a JFT-2 tokamak plasma heated by ion cyclotron wave and neutral beam injection. It was found that the energy spectrum could be determined with sufficient accuracy, and the ion temperature obtained from the energy spectrum was in good agreement with the ion temperature deduced from the Doppler broadening of the impurity ion line.Keywords
This publication has 5 references indexed in Scilit:
- Determination of Ion Temperature Distributions from Doppler Broadenings of Impurity Lines in the Vacuum UV Region for Tokamak DischargesJournal of the Physics Society Japan, 1979
- The Local Counting Efficiency of a Ceratron Multiplier for Plasma Diagnoses on Charge-Exchanged Fast AtomsJapanese Journal of Applied Physics, 1977
- A 10-Channel Neutral Particle Energy Analyser for Measurements of Ion Temperatures on Tokamak PlasmasJapanese Journal of Applied Physics, 1977
- Electron Stripping of He and H in Helium, Hydrogen and Nitrogen Gases from 0.2 keV to 5.0 keVJournal of the Physics Society Japan, 1976