Theoretical Maximal Precision for Mass-to-Charge Ratio, Amplitude, and Width Measurements in Ion-Counting Mass Analyzers
- 13 April 2000
- journal article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 72 (10) , 2256-2260
- https://doi.org/10.1021/ac991256u
Abstract
A theory previously developed for spectra with detector-limited (i.e., signal-independent) Gaussian-distributed noise is applied to calculate the maximal precision with which mass spectral peak parameters (mass-to-charge ratio, amplitude, width) can be determined from a discrete spectrum with source-limited Poisson-distributed noise. The precision depends in a calculable way upon the peak shape, signal-to-noise ratio, and number of data points per peak width. Those dependencies are tested by analysis of simulated data. The theory provides estimates for the precision of a repeated experiment, based on data from a single discrete mass spectrum whose parameters are extracted by a least-squares fit to a specified line shape. The relevance of the predictions to present and potential time-of-flight performance is discussed.Keywords
This publication has 4 references indexed in Scilit:
- Time-Domain (Interferogram) and Frequency-Domain (Absorption-Mode and Magnitude-Mode) Noise and Precision in Fourier Transform SpectrometryApplied Spectroscopy, 1990
- Probability and StatisticsPublished by Springer Nature ,1989
- Effect of signal-to-noise ratio and number of data points upon precision in measurement of peak amplitude, position and width in fourier transform spectrometryChemometrics and Intelligent Laboratory Systems, 1986
- Fourier, Hadamard, and Hilbert Transforms in ChemistryPublished by Springer Nature ,1982