Specific and Sensitive Measurement of FK506 and Its Metabolites in Blood and Urine of Liver-Graft Recipients
- 1 October 1992
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 38 (10) , 2025-2032
- https://doi.org/10.1093/clinchem/38.10.2025
Abstract
A specific and sensitive assay for quantifying the immunosuppressant FK506 and its metabolites in blood and urine was developed. 32-O-Acetyl FK506 was synthesized and used as internal standard. FK506 and its metabolites were purified from the samples by solid-liquid extraction and were injected into a high-performance liquid chromatographic (HPLC) system linked to a mass spectrometer (MS) by particle-beam interface. The FK506 derivatives were separated from interfering material by use of a 100 x 4 mm C8 analytical column and water/acetonitrile or water/methanol gradient elution; they were detected by negative chemical ionization with methane as reagent gas. The limit of detection was 25 pg in a standard solution, and the limit of quantification in blood was 250 pg (extracted from 1 mL of blood). The CV was 11.3% at 5 ng, and no interferences with other drugs were found.Keywords
This publication has 0 references indexed in Scilit: