First High-Performance Liquid Chromatography Assay of Formoterol Concentrations in the Low-Picogram-per-Milliliter Range
- 1 April 1994
- journal article
- Published by Wolters Kluwer Health in Therapeutic Drug Monitoring
- Vol. 16 (2) , 196-199
- https://doi.org/10.1097/00007691-199404000-00015
Abstract
A method for the assay of plasma concentrations of the long-acting p2-adrenoceptor agonist, formoterol is described. This method, in which high-performance liquid chromotography is used with electrochemical detection, enables, for the first time, pharmacokinetic and pharmacodynamic research with this drug. After column extraction of plasma samples, the eluent is injected on the chromatographic system. Retention times of formoterol and the internal standard, bromo-formoterol, were 5.6 and 10 min, respectively. Our results show that this is a sensitive and reproducible method with a very low limit of detection (20 pg/ml), which creates the possibility of measuring concentrations in a range achievable in humans.Keywords
This publication has 0 references indexed in Scilit: