Spatial Resolution of the PEP-4 Time Projection Chamber
- 1 January 1983
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 30 (1) , 76-81
- https://doi.org/10.1109/tns.1983.4332223
Abstract
The spatial resolution and response of the segmented cathode pads of the PEP-4 TPC have been measured with data taken at 8.5 atmospheres of 80% Argon-20% Methane gas with a 4kG magnetic field. The dependence of the spatial resolution and pad response on drift distance and track-anode crossing angle is presented.Keywords
This publication has 5 references indexed in Scilit:
- Gated time projection chamberNuclear Instruments and Methods in Physics Research, 1983
- Measurement of Ionization Loss in the Relativistic Rise Region with the Time Projection ChamberIEEE Transactions on Nuclear Science, 1983
- Performance of the Signal Processing System for the Time Projection ChamberIEEE Transactions on Nuclear Science, 1983
- Performance of a time-projection chamberNuclear Instruments and Methods, 1979