Complications of Semiflexible, Closed-Loop Anterior Chamber Intraocular Lenses
- 1 January 1987
- journal article
- research article
- Published by American Medical Association (AMA) in Archives of Ophthalmology (1950)
- Vol. 105 (1) , 52-57
- https://doi.org/10.1001/archopht.1987.01060010058031
Abstract
• Anterior chamber intraocular lenses (IOLs) provide good visual rehabilitation of aphakia in a majority of patients. An increasing number of eyes with closedloop, semiflexible anterior chamber implants are now being seen with decreased vision due to corneal edema and persistent cystoid macular edema. We evaluated 53 such eyes in 52 patients who presented with decreased vision one to 51 months (average, 15 months) after lens implantation. The most frequent lens implanted was the Surgidev Style 10 Leiske IOL. Surgery for IOL removal or exchange with or without penetrating keratoplasty was performed in 34 (64%) of 53 eyes; visual acuity recovery ranged from 20/20 to hand motions (average, 20/80). Despite clear corneal grafts in 24 (92%) of the 26 eyes that underwent corneal transplantation, visual acuity of 20/ 40 or better was obtained in only nine eyes (26%). Based on the intractable inflammatory sequelae associated with anterior chamber closed-loop IOLs, we strongly urge discontinuation of their use.This publication has 12 references indexed in Scilit:
- Closed-Loop Anterior Chamber LensesArchives of Ophthalmology (1950), 1987
- Effect of Intraocular Lens Fixation on the Blood-Aqueous BarrierAmerican Journal of Ophthalmology, 1984
- A comparative study of the 91Z and other anterior chamber intraocular lensesAmerican Intra-Ocular Implant Society Journal, 1984
- Five cases of severe anterior chamber lens implant complicationsAmerican Intra-Ocular Implant Society Journal, 1984
- Complications while removing the IOLAB 91Z lens for the UGH-UGH+ syndromeAmerican Intra-Ocular Implant Society Journal, 1984
- Late postoperative hemorrhage following intracapsular cataract extraction with the IOLAB 91Z anterior chamber lensAmerican Intra-Ocular Implant Society Journal, 1983
- Loop Precipitations in the 91Z LensAmerican Intra-Ocular Implant Society Journal, 1983
- UGH Syndrome with the 91Z LensAmerican Intra-Ocular Implant Society Journal, 1983
- Lens-induced complications with anterior chamber dens implants: A comparison with iris supported and posterior chamber lensesAmerican Intra-Ocular Implant Society Journal, 1983
- Iris Erosion with the Leiske LensAmerican Intra-Ocular Implant Society Journal, 1982