Thermocurrents in PTFE prepared for use in dosimetry

Abstract
Thermocurrent studies have been made of PTFE and FEP polymers as prepared for radiation dosimetry applications. Samples are processed under controlled annealing conditions. Three cooling rates are employed: (a) 16 degrees C h-1, (b) 180 degrees C h-1 and (c) quenched, to produce samples of known morphology and sensitivities: (a) 250 pC rad1 cm-2, (b) 170 pC rad-1 cm-2 and (c) 140 pC rad-1 cm-2. Unannealed samples have a sensitivity of 75 pC rad-1 cm-2. Results show that initial small doses of radiation cause a 100% change in the magnitude of the TC spectra and a 30-40 degrees C shift in the peak temperature. The authors postulate annihilation of low-energy traps and suggest a correlation between carrier path length and morphological size. A spurious signal equivalent to 15 mrad accumulates during storage. Dose response curves indicate the usefulness of PTFE and FEP for dosimetry.