Interactive MR imaging-guided biopsy and aspiration with a modified clinical C-arm system.

Abstract
This investigation was performed to test the hypotheses that interactive guidance of MR image acquisition during needle-directed procedures using a clinical 0.2-T C-arm open MR imaging system integrated with a frameless optically linked stereotaxy system is feasible, and that procedure times can be sufficiently short to be well tolerated by the patient.One hundred six MR-guided procedures were performed in 86 patients (ranging in age from 5 months to 88 years) using a clinical C-arm imaging system supplemented with an in-room RF-shielded liquid crystal display monitor, a frameless stereotaxy system, rapid gradient-echo sequences for needle guidance, and MR-compatible monitoring and surgical lighting equipment. We performed 50 biopsies and aspirations of the head and neck in 37 patients, 23 biopsies of musculoskeletal lesions in 22 patients, 16 biopsies of abdominal sites in 10 patients, six biopsies of the thoracolumbar spine or sacrum in six patients, and 11 shoulder joint injections for MR arthrography ...

This publication has 0 references indexed in Scilit: