Photoelectric recording of mechanical responses of cardiac myocytes

Abstract
A method to monitor contraction of isolated myocytes by transmicroscopic photometry is illustrated. Two photodiodes are mounted inside an inverse microscope used for visual control of a cell. Illumination of one diode varies in proportion to changes in cell length. The contraction signal is amplified in a comparator circuit. Spatial resolution of the device is in the order of 1 μm which corresponds to about 5% of cell shortening in the fully activated state of contraction. The method was tested on isolated myocytes from guinea-pig ventricle. Optical records of contraction in response to action potentials or during voltage clamp compare well with the contractile behaviour of multicellular preparations.