Dynamic-range compression in surface-coil MRI
- 1 August 1986
- journal article
- research article
- Published by American Roentgen Ray Society in American Journal of Roentgenology
- Vol. 147 (2) , 379-382
- https://doi.org/10.2214/ajr.147.2.379
Abstract
The large dynamic range in signal intensity present in MR surface-coil images makes proper windowing and photography difficult. By removing the low-spatial frequency information caused by variations in surface-coil field intensity from the high-spatial frequency information containing the image data, a considerable compression of this dynamic range of signal intensity is possible. To accomplish this, a technique was implemented on a digital computer for use with MR surface-coil image data. The compression was done as a postprocessing option after the patient scan had been completed and therefore did not alter actual scan times. Although the image signal-to-noise was not altered, the ease of photography for most images was improved. Thus, digital dynamic-range compression is a practical technique to aid in surface-coil MRI studies.This publication has 4 references indexed in Scilit:
- Solenoid surface coils in magnetic resonance imagingAmerican Journal of Roentgenology, 1986
- HIGH-FIELD SURFACE-COIL MR IMAGING OF LOCALIZED ANATOMY1985
- Surface Coil Magnetic Resonance ImagingJournal of Computer Assisted Tomography, 1984
- Mapping of metabolites in whole animals by 31P NMR using surface coilsNature, 1980