Temporal and spatial analysis of fields generated by eddy currents in superconducting magnets: Optimization of corrections and quantitative characterization of magnet/gradient systems
Open Access
- 1 August 1991
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 20 (2) , 268-284
- https://doi.org/10.1002/mrm.1910200209
Abstract
We propose methods for the spatial and temporal characterization of time‐dependent magnetic fields generated by eddy currents after switching gradients. For an on‐line determination of the temporal variations of the fields, we extract two terms from the unresolved signal of an extended sample, describing the time evolution of a frequency shift γΔBz (t) and of a decay constant k(t). This procedure allows us to optimize interactively the multiexponential pre‐emphasis as well as any spectral volume selection method with respect to eddy currents, Additionally, we suggest an imaging sequence which allows us to determine the spatial distribution of eddy current fields at a chosen time‐point after any gradient sequence to be tested. Expansion of these eddy currents fields into spherical harmonic functions proves the existence of a higher order terms, which cannot be corrected by a standard pre‐emphasis device, where time constants and amplitudes are adjusted on the X, Y, Z and Z0 coils. The proposed numerical analysis gives a tool to characterize any magnet/ gradient system quantitatively with respect to eddy current performance. © 1991 Academic Press, Inc.Keywords
This publication has 9 references indexed in Scilit:
- Application of interlaced orientation slices in orthogonal planes for system characterization and as a basis for standardized brain sections in multiple‐oblique positionsMagnetic Resonance in Medicine, 1990
- Correction of distortions due to the pulsed magnetic field gradient-induced shift in B0 field by postprocessingMagnetic Resonance in Medicine, 1989
- Combined application of MR imaging and spectroscopy in neonates and children: installation and operation of a 2.35-T system in a clinical setting.Radiology, 1988
- The correction of transient B0 field shifts following the application of pulsed gradients by phase correction in the time domainJournal of Magnetic Resonance (1969), 1986
- Image-selected in Vivo spectroscopy (ISIS). A new technique for spatially selective nmr spectroscopyJournal of Magnetic Resonance (1969), 1986
- Active magnetic screening of gradient coils in NMR imagingJournal of Magnetic Resonance (1969), 1986
- Magnet systems used in medical NMRComputerized Radiology, 1984
- Magnet field profiling: Analysis and correcting coil designMagnetic Resonance in Medicine, 1984