Magnetic Resonance Imaging with Implanted Neurostimulators: An In Vitro and In Vivo Study
- 1 January 1999
- journal article
- research article
- Published by Wolters Kluwer Health in Neurosurgery
- Vol. 44 (1) , 118-125
- https://doi.org/10.1097/00006123-199901000-00073
Abstract
The goal was to assess the safety of magnetic resonance imaging (MRI) with implanted neurostimulators, in an in vitro and in vivo study. Two different implantable pulse generators (IPGs) (ITREL II and 3; Medtronic, Minneapolis, MN) and different leads (separately and connected to an IPG) were tested in three different MRI scanners (0.2, 0.25, and 1.5 T). Measurements of the induced voltages (using an external oscilloscope) and the induced heat (using an infrared camera) were performed in an in vitro study. Finally, 38 patients with implanted neurostimulator systems (leads and IPGs) underwent MRI in 50 examinations, with continuous monitoring by a physician with uninterrupted visual and vocal contact with the patient. Twenty-five patients were studied prospectively, with documented printouts of the parameter settings before and after MRI. An induced voltage of 2.4 to 5.5 V was measured in the experimental configuration with a lead connected to an IPG. The voltage was higher with the leads alone, compared with the leads connected to the IPG, and was dependent on the MRI scanner, the sequences, and the type of lead. No heat induction was observed in any part of the hardware. No change of pulse shape or change of IPG parameters was observed during MRI. No adverse effects occurred in patients with chronically implanted deep brain leads connected to an IPG. MRI can be safely performed in patients with implanted neurostimulation systems with the tested deep brain leads connected to an IPG (ITREL II and 3), with running parameters. No heat induction was detected, and the experimentally measured induced voltage did not seem to harm the patients. Only the reed switch of the IPGs was activated; the other parameters remained unchanged. Further investigations must be performed to study the local electrical effects in larger plate electrodes; these effects might cause slight discomfort. There is no danger with any type of electrode during MRI examinations if the electrodes lie outside the region of interest. These observations are restricted to the tested devices. A conscientious estimation of the risks and benefits of MRI for patients with implanted devices is recommended. If the type of device is not known to the examiner, MRI should still be considered to be contraindicated.Keywords
This publication has 10 references indexed in Scilit:
- MRT bei Patienten mit Herzschrittmachern: In-vitro- und In-vivo-Evaluierung bei 0,5 TeslaRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 1998
- Aneurysm clips: magnetic quantification and magnetic resonance imaging safetyJournal of Neurosurgery, 1997
- Effects of magnetic resonance imaging on cardiac pacemakers and electrodesAmerican Heart Journal, 1997
- Magnetic resonance imaging and spinal cord stimulation systemsPAIN®, 1997
- Effects of Nuclear Magnetic Resonance Imaging on Cardiac PacemakersPacing and Clinical Electrophysiology, 1995
- The Effect of Magnetic Resonance Imagers on Implanted NeurostimulatorsPacing and Clinical Electrophysiology, 1992
- Effect of 1.5 tesla nuclear magnetic resonance imaging scanner on implanted permanent pacemakersJournal of the American College of Cardiology, 1987
- The Effects of Nuclear Magnetic Resonance Imagers on External and Implantable Pulse GeneratorsPacing and Clinical Electrophysiology, 1984
- Potential hazards and artifacts of ferromagnetic and nonferromagnetic surgical and dental materials and devices in nuclear magnetic resonance imaging.Radiology, 1983
- Which elements are excited in electrical stimulation of mammalian central nervous system: A reviewBrain Research, 1975