Abstract
Summary: Flow cytometry, the method of choice for determining subset cell populations, requires sophisticated instrumentation and highly trained operators. An alternative method for subpopulation determinations is presented that is fast, simple, and readily interpreted for absolute subset counts. This method uses antibody labeled microspheres added to whole blood and after lysing, read on a hemocytometer using an ordinary light microscope. The total time of testing is less than 10 min with results comparable to that obtained using a flow cytometer and a hematology analyzer to calculate absolute subpopulation values. This method is demonstrated for CD4+ lymphocytes which have considerable value in following HIV-infected individuals.Keywords
This publication has 5 references indexed in Scilit:
- Laboratory control values for CD4 and CD8 T lymphocytes. Implications for HIV-1 diagnosisClinical and Experimental Immunology, 1992
- Incidence of HIV infection in monocyte subpopulations characterized by CD4 and HLA-DR surface densityAIDS, 1992
- Measuring percent lymphocytes by flow cytometry to calculate absolute lymphocyte subset counts for HIV + specimensCytometry, 1992
- The natural history of HIV-1 infectionAIDS, 1991
- The Prognostic Value of Cellular and Serologic Markers in Infection with Human Immunodeficiency Virus Type 1New England Journal of Medicine, 1990