Abstract
The analytical performance of the NucliSens HIV-1 QT assay, a highly sensitive test based on nucleic acid sequence-based amplification technology, was evaluated in a multicenter trial. Assay specificity was evaluated with 502 plasma (EDTA) specimens from human immunodeficiency virus type 1 (HIV-1)-seronegative volunteer donors. No HIV-1 RNA was reported in any of the donor specimens. Analytical sensitivity and reproducibility were estimated with panels prepared from a high-titer well-characterized HIV-1 RNA stock (5.84 × 10 8 RNA copies/ml). The assay's dynamic range was linear from 10 6 to 10 1 HIV-1 RNA copies, with a lower detectable limit of 25 copies/ml and a 95% detection rate of 176 copies/ml. Sensitivity of the assay to detect HIV-1 RNA in clinical specimens from patients ( n = 101) and in commercially available or prepared panels ( n = 24) was compared with NASBA HIV-1 RNA QT (an earlier version of NucliSens HIV-1 QT) and with the Food and Drug Administration-approved standard and ultrasensitive AMPLICOR HIV-1 MONITOR, version 1.0, assays. Detection of HIV-1 RNA was reproducible over a 5-log range (mean standard deviation = 0.15 log). The NucliSens and the standard AMPLICOR assays were equivalent in detection of HIV-1 RNA (concentration, 10 3 to 10 5 copies/ml) in 57 clinical specimens. The NucliSens assay was more sensitive in detecting HIV-1 RNA at lower concentrations (≤10 2 copies/ml) (44 of 44) than either the standard AMPLICOR test (12 of 19) or the NASBA assay (10 of 25). A 25% increase in HIV-1 RNA detection frequency with panels was observed with the NucliSens assay (23 of 24) compared with the standard AMPLICOR test (17 of 24). The new assay was highly specific and demonstrated good sensitivity with a broad linear dynamic range.Keywords
This publication has 21 references indexed in Scilit:
- HIV-1 Viral Load TestingLaboratory Medicine, 2001
- The contribution of assay variation and biological variation to the total variability of plasma HIV-1 RNA measurementsAIDS, 1999
- Comparison of three commercial assays for the quantification of HIV-1 RNA in plasma from individuals infected with different HIV-1 subtypesJournal of Clinical Virology, 1999
- Comparison of Three Different Commercial Methods for Measuring Plasma Viraemia in Patients Infected with Non-B HIV-1 SubtypesEuropean Journal of Clinical Microbiology & Infectious Diseases, 1999
- Impact of HIV Type 1 Subtype Variation on Viral RNA QuantitationAIDS Research and Human Retroviruses, 1999
- HIV-1 RNA Detection in Plasma for the Diagnosis of Infection in NeonatesJAIDS Journal of Acquired Immune Deficiency Syndromes, 1997
- Viral Load and Disease Progression in Infants Infected with Human Immunodeficiency Virus Type 1New England Journal of Medicine, 1997
- HIV-1 Dynamics in Vivo: Virion Clearance Rate, Infected Cell Life-Span, and Viral Generation TimeScience, 1996
- Rapid turnover of plasma virions and CD4 lymphocytes in HIV-1 infectionNature, 1995
- Viral dynamics in human immunodeficiency virus type 1 infectionNature, 1995