Evaluation of Glucocard Memory 2 and Accutrend® Sensor Blood Glucose Meters
- 30 January 1998
- journal article
- research article
- Published by Walter de Gruyter GmbH in cclm
- Vol. 36 (1) , 47-52
- https://doi.org/10.1515/cclm.1998.009
Abstract
The performance and practicability of 2 blood glucose meters (Glucocard Memory 2 and Accutrend(R) sensor) were evaluated. Both glucose meters produced acceptably precise results in the hyper-and normoglycaemic concentration ranges. In the hypoglycaemic concentration range, the imprecision of Accutrend(R) sensor was much higher than recommended by the American Diabetes Association. Within-run coefficients of variation for Glucocard Memory 2 were 6.3 %, 3.9 % and 2.4 % at glucose concentrations of 1.7 mmol/l, 5.8 mmol/l and 11.7 mmol/l, respectively; for Accutrend(R) sensor these were 15.2 %, 5.0 % and 1.2% at respective concentrations of 0.9 mmol/l, 4.2 mmol/l and 19.6 mmol/l. Between-day coefficients of variation for Glucocard Memory 2 were 4.8 % and 3.5 % at glucose concentrations of 3.9 mmol/l and 17.2 mmol/l, respectively and for Accutrend(R) sensor they were 3.8 % and 2.9 % at glucose concentrations of 3.8 mmol/land 18.7 mmol/l, respectively. Results were linear over a range of 1.6 mmol/l - 29.7 mmol/l for Glucocard Memory 2 and 1.6 mmol/l - 33.3 mmol/l for Accutrend(R) sensor. Results of both blood glucose meters correlated closely with the hexokinase/glucose-6-phosphate dehydrogenase laboratory method. Ninety-eight percent of both Glucocard Memory 2 and Accutrend(R) sensor results were within 20 % of the comparison method values. Ninety-three percent of the Glucocard Memory 2 and 96 % of the Accutrend(R) sensor results were within 15 % of the comparison method results. An inverse relation between the glucose readings and haematocrit values was observed for both blood glucose meters in the hyperglycaemic range and this effect was more pronounced for Accutrend(R) sensor. In the normo-and hypoglycaemic ranges the effect was insignificant and absent, respectively. Minimum sample volume for Glucocard Memory 2 was 3 mu l and for Accutrend(R) sensor it was 9 mu l. Lower sample volumes gave erroneous results. Presenting more than the required volume had no effect on results.Keywords
This publication has 7 references indexed in Scilit:
- The Effect of Intensive Treatment of Diabetes on the Development and Progression of Long-Term Complications in Insulin-Dependent Diabetes MellitusNew England Journal of Medicine, 1993
- The Effect of Haematocrit on Reagent Strip Tests for GlucoseDiabetic Medicine, 1991
- The Quality of Self‐monitoring of Blood GlucoseDiabetic Medicine, 1990
- Effect of Hematocrit Concentration on Blood Glucose Value Determined on Glucometer IIDiabetes Care, 1988
- Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood GlucoseDiabetes Care, 1987
- Gestational Diabetes MellitusDiabetes Care, 1986
- Control of Maturity-Onset Diabetes by Monitoring Fasting Blood Glucose and Body WeightDiabetes Care, 1980