Improving wedged field dose distributions

Abstract
Dose profiles produced by wedge filters in the non-wedged direction can exhibit a 7% or greater dose reduction at the outer ends of the field compared with open field profiles. However, many planning systems use open field profiles to model wedged dose distributions. In the present work, wedges have been modified to reproduce open field profile shapes. This modification involved removing varying thicknesses of the wedge using a simple milling machine. The wedge thickness was calculated using the assumption that dose is proportional to primary collision kerma. The discrepancies in dose between wedged field and open field profile shapes of up to 7% were reduced to less than 3% with the modifications, even for varying depths and off-axis distances. The necessary measurements are simple to perform, and hence this technique could be applied to improve wedged field dose distributions in other radiotherapy departments.